IEC 60601-1-2

From RadiWiki
Revision as of 13:55, 4 February 2022 by Joro (talk | contribs) (Created page with "The EN/IEC 60601-1-2 standard describes how EMC tests are performed on medical electrical equipment The exact title of this standard is: "IEC 60601-1-2: Medical elect...")
(diff) ← Older revision | Latest revision (diff) | Newer revision → (diff)
Jump to navigation Jump to search

The EN/IEC 60601-1-2 standard describes how EMC tests are performed on medical electrical equipment

The exact title of this standard is: "IEC 60601-1-2: Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic disturbances - Requirements and tests"

This review is based on the IEC 60601-1-2:2014/A1:2020 version.

For a lot of the requirements, this standard is referring to other IEC standards. Some of these referenced standards are:

In section 8.11, it is described that the magnetic field close proximity test as described in IEC 61000-4-39 only needs to be performed on 3 frequencies (30 kHz, 134.2 kHz and 13.56 MHz)

Links[edit]