RadiSenseDeviceDriverConfiguration

From RadiWiki
Revision as of 11:34, 8 March 2021 by Joro (talk | contribs)
Jump to navigation Jump to search
Information.png
Note: This page is used to show the configuration of the DARE!! Instruments RadiSense device drivers. This page can now easily be embedded on all relevant device driver pages.

The following tabs are available in the advanced configuration of the RadiSenseDeviceDriverConfiguration:

GPIB DeviceStream Configuration.png

On the Communication tab, the desired communication method can be selected and configured. Depending on the selected method, additional relevant settings are shown and can be configured.


ScreenElementDescription.svg Communication Streams Selects the medium or method that should be used to communicate with the device. Depending on the capabilities of the device this can be one or more of:

See the Communication Settings in Chapter 15, on how to configure each of these methods.


FieldProbePanel.png


ScreenElementDescription.svg Pre Wait Time Wait time before starting the measurement-loop. Can be used to wait for a signal generator, and or amplifier to stabilize.
ScreenElementDescription.svg Measure The minimum amount of measurements in the measure loop. Continues to loop even if field strength level is stable.
ScreenElementDescription.svg Wait Time The wait time between each measurement is taken, during the measurement-loop.
ScreenElementDescription.svg Max. Difference If the measured field strength between each measurement in the measurement-loop, is less than this value. The measured field strength is stable. The measurement-loop will then stop.
ScreenElementDescription.svg Max. Measure The maximum amount of measure times, for the measurement-loop


According to these settings, RadiMation® will perform a minimum of 3 field strength measurements with a wait time between the measurements of 75ms. If all those 3 measurements are within 0.5 V/m, the field strength measurement is assumed to be stable, and RadiMation® will continue to average of the last 3 measurements. If the difference between the measurements is larger than the specified 0.5 V/m, an additional field strength measurement will be performed, and the last 3 samples will be compared to the 0.5 V/m difference again. At most 10 measurements will be performed if the difference between the 3 samples will not be within the specified 0.5 V/m.

Both for the leveling, and the monitoring these settings can be set individually. The leveling is used, whenever RadiMation® is measuring the field strength, to regulate the signal generator. If the field sensor is only used as input, without regulating on it, then the monitoring settings are used.

More detailed information on how to achieve a fast and stable field strength measurement are available in Application Note 106: Powermeter parameter configuration.

ConfigurableFrequencyRangeTab.png

The frequency range of the RadiSenseDeviceDriverConfiguration as provided by the manufacturer is shown and selected as default. It is possible to overrule these frequencies and to manual adjust the allowed frequency range of the RadiSenseDeviceDriverConfiguration.

ScreenElementDescription.svg Use default start frequency If the checkbox is checked, the default start frequency will be used as the lowest usable frequency in a test for this device.
ScreenElementDescription.svg Custom start frequency If the Use default start frequency checkbox is unchecked, another start frequency (expressed in MHz) can be specified. The customized start frequency will then be used as the lowest usable frequency in a test for this device. The customized frequency can be a limitation or an extension of the default start frequency.
ScreenElementDescription.svg Use default stop frequency If the checkbox is checked, the default stop frequency will be used as the highest usable frequency in a test for this device.
ScreenElementDescription.svg Custom stop frequency If the Use default stop frequency checkbox is unchecked, another stop frequency (expressed in MHz) can be specified. The customized stop frequency will then be used as the highest usable frequency in a test for this device. The customized frequency can be a limitation or an extension of the default stop frequency.


Specifying a different frequency range can be useful if for example:

  • A device (like a coupler, antenna, injection device, cable, etc...) is still useable (but out of specification) outside the standard suggested frequency range.
  • An external mixer is used to measure an extended frequency range.
  • An up- or down-convertor is used to shift the frequency range.
  • A newer model of a device is present that has an extended frequency range, and still uses the same remote control commands.

Be careful changing these setting as RadiMation® is no longer able to verify if the RadiSenseDeviceDriverConfiguration is used outside frequency range that is specified by the manufacturer. This may result to serious damage of your measurement device.

DeviceDriverRadiCentreTab.png

There are several methods in which the RadiSenseDeviceDriverConfiguration can be connected to the PC. The RadiSenseDeviceDriverConfiguration can for example be connected directly to the PC, or through a plugincard in a RadiCentre 1 or a multi-slot RadiCentre. More information on how to use the RadiCentre is present in the product manual of the RadiCentre. The settings on the RadiCentre tab can be used to specify how the RadiSenseDeviceDriverConfiguration is connected.

When the RadiSenseDeviceDriverConfiguration is connected to a multi-slot RadiCentre, also the slot number in which the plugincard is fitted, should be specified.


ScreenElementDescription.svg Directly to PC Specifies that the RadiSenseDeviceDriverConfiguration is connected directly to the PC, without the use of a RadiCentre.
ScreenElementDescription.svg RadiCentre 1 Specifies that the RadiSenseDeviceDriverConfiguration is connected to a plugincard that is fitted in a RadiCentre 1.
ScreenElementDescription.svg RadiCentre multi-slot Specifies that the RadiSenseDeviceDriverConfiguration is connected to a plugincard that is fitted in a multi-slot RadiCentre that also can be manually controlled using the display on the front of the RadiCentre. The slot number in which the plugin card is mounted can also be selected.
ScreenElementDescription.svg RadiCentre multi-slot Ultra Specifies that the RadiSenseDeviceDriverConfiguration is connected to a plugincard that is fitted in a multi-slot RadiCentre Ultra. This RadiCentre Ultra has no display, and can only be used when remote controlled.


The settings on this RadiCentre tab, only specifies if and which RadiCentre is used. The communication settings on the Communication tab specify the connection settings to communication with the measurement device.

The configuration on the Communication tab should be configured depending on the outgoing connection from the PC. Thus if Directly to PC is selected, the communication settings should be configured as what is actually used between the PC and the RadiSenseDeviceDriverConfiguration itself. If RadiCentre 1, RadiCentre multi-slot or RadiCentre multi-slot Ultra is selected, the communication settings should be configured as what is actually used between the PC and the RadiCentre. In the configuration where a RadiPower is connected with USB to a multi-slot RadiCentre, which is connected by GPIB to the PC, the communication settings of the RadiPower device driver should thus be GPIB.

If the RadiSenseDeviceDriverConfiguration is for example connected to slot 3 of a multi-slot RadiCentre which has IP address 192.168.178.95 and is connected by LAN to the PC, the following settings should be configured in the device driver of the RadiSenseDeviceDriverConfiguration:

  • RadiCentre tab
    • Connection type: RadiCentre multi-slot, using as slot number: 3
  • Communication tab
    • Communication Stream: VISA
    • LAN: 192.168.178.95

DeviceDriverRadiCentreDeviceSoftwareUpdateTab

DeviceDriverRadiSenseTab.png

The RadiSense tab can be used to configure the specific settings of the RadiSenseDeviceDriverConfiguration.

ScreenElementDescription.svg Direction If the field strength of a specific single axis or an isotropic value should be measured.
ScreenElementDescription.svg Band selection Allows if the Low frequency band, High frequency band, or both frequency bands should be used for the electrical field strength measurement.
ScreenElementDescription.svg RadiSense Firmware Version Allows to select the firmware version that is present in the RadiSense probe. This information is used to used version specific commands to control the RadiSense and which baudrate to use for serial communication (version 4.6.0 and below use 9k6, versions above 4.6.0 use 19k2).
ScreenElementDescription.svg Filter Setting The filter that should be used during the measurements.
ScreenElementDescription.svg Automatically turn laser on If this option is checked, the driver will automatically activate the laser when the device is being used.