GraphExport.CALFile.MinimumField

From RadiWiki
Revision as of 09:58, 26 April 2024 by Joro (talk | contribs) (Created page with "This advanced option specifies the minimum electrical field that should at least be achieved to allow the export of the field strength graph to a...")
(diff) ← Older revision | Latest revision (diff) | Newer revision → (diff)
Jump to navigation Jump to search

This advanced option specifies the minimum electrical field that should at least be achieved to allow the export of the field strength graph to a radiated immunity calibration file.

If on one or more frequencies the field strength is below 1.0 V/m, the field strength cannot be exported to a radiated immunity graph, and the validation error is shown: "The electrical field graph contains data points that are below 1 V/m. The electrical field graph should at least be 1 V/m on all data points."

By default this value is 1.0 V/m. If for some reason it is required to allow export of an electrical field graph to a radiated immunity graph, to also contain values lower than 1.0 V/m, it is possible to set this value to 0.0 V/m, which will allow to export to the calibration file, even if the field strength was 0.0 V/m at one or more frequencies.

Warning-48px.png
Warning: Any change and/or modification to the advanced options, can result in unexpected results and/or incorrect measurement results. Changes in any advanced option should be verified and checked very carefully!