RadiMation Error: 805330289

From RadiWiki
Revision as of 13:57, 3 September 2024 by Roge (talk | contribs) (→‎Solution)
(diff) ← Older revision | Latest revision (diff) | Newer revision → (diff)
Jump to navigation Jump to search

Polarizer has stopped before reaching the desired destination.

The RadiField polarizer has started to rotate the RadiField to the desired polarization however it couldn't achieve this operation, and has stopped before the desired destination is reached.

The RadiMation® software can raise this error during the initialization and check of the device, even if the polarization of the RadiField can be changed from the RadiCentre display without problems. In that case the calibration of the polarizer is to close to the end of the movement for the opposite polarization.

Solution

  • Check if the RadiField polarizer is working correctly when controlled manually.
  • Calibrate the horizontal and vertical polarization of the RadiField again in the RadiCentre configuration.
  • If the calibration was performed but still raises this error in RadiMation, make sure to put the polarization to horizontal or vertical manually with the RadiCentre display.