Revision as of 12:33, 10 September 2025 by Joro (talk | contribs) (Created page with "= The Average detector can only be used if it is the only active detector. Disable other detectors = For the Rohde & Schwarz spectrum analyser, multiple trace detectors are activated, one of them being the average detector. The average detector can however only be used during the prescan if it is the only detector. == Solution == * Disable the average detector during the prescan. * Disable all the other detectors (and only enable the average detector) during the prescan...")
(diff) ← Older revision | Latest revision (diff) | Newer revision → (diff)
Jump to navigation Jump to search

The Average detector can only be used if it is the only active detector. Disable other detectors

For the Rohde & Schwarz spectrum analyser, multiple trace detectors are activated, one of them being the average detector. The average detector can however only be used during the prescan if it is the only detector.

Solution

  • Disable the average detector during the prescan.
  • Disable all the other detectors (and only enable the average detector) during the prescan.