"<device>" can not be used for this test, valid frequency range from <lower frequency> to <upper frequency>.

During the initialization of the testsite, the allowed frequency range of the used equipment ("<device>") is checked. This error is raised if it is detected that the device is limited in the frequency range, and cannot be used for the complete frequency range of the test. Because of this problem the test cannot be started.

Solution

  • Reduce the frequency range of the test. For example by increasing the start frequency of the test, or decreasing the stop frequency of the test
  • Extend the allowed frequency range of the device.
  • Check the minimum and maximum values in the frequency column of the correction files attached to the device.