RadiMation Error: 536871941
Jump to navigation
Jump to search
Start Frequency is too low for the selected equipment. Usable frequency range of '<device>' is from <start-frequency> until <stop-frequency>
During the initialization of the testsite, the allowed frequency range of the used equipment ("<device>") is checked. This error is raised if it is detected that the device is limited in the frequency range, and cannot be used for the complete frequency range of the test. Because of this problem the test cannot be started.
Solution
- Select another device in the test-site replacing the device that is limiting the frequency range
- Reduce the frequency range of the test. For example by increasing the start frequency of the test, or decreasing the stop frequency of the test
- Extend the allowed frequency range of the device.
- Check the minimum and maximum values in the frequency column of the correction files attached to the device.
- To find out the device that is causing the problems, some devices can be replaced by 'Virtual' equipment, until the error is not shown anymore. All Virtual equipment is usable from 1 Hz to 40 GHz.