RadiMation Error: 805309219

From RadiWiki
Revision as of 10:15, 5 May 2017 by Sast (talk | contribs) (→‎Solution)
(diff) ← Older revision | Latest revision (diff) | Newer revision → (diff)
Jump to navigation Jump to search
The printable version is no longer supported and may have rendering errors. Please update your browser bookmarks and please use the default browser print function instead.

The calibration of the device caused a timeout: no confirmation could be retrieved within the maximum timeout[edit]

When the calibration setting of the device is enabled the driver will try to have the device perform a (short) calibration e.g. before the final measurement.

The first time the calibration is being activated this can take up to 90 seconds! This can cause (GPIB) timeouts, as the driver expects a confirmation after the calibration is finished. It is strongly advised to perform the first calibration of the day manually at the device itself. When it is not desirable that a calibration is performed please disable this function.

Solution[edit]

  • Perform a manually calibration before remote controlling the device.
  • Disable the automatic calibration function during the test in the Advanced Device Driver settings.