RadiMation Error: 536871966

From RadiWiki
Revision as of 14:49, 19 May 2021 by Joro (talk | contribs)
Jump to navigation Jump to search

There is no average probe selected[edit]

This error happens when you are doing a Radiated Immunity calibration and do not have at least one averaging probe selected. For a one-point radiated immunity calibration at least one electrical field strength probe should be selected to be included in the calculation of the average field.

Also see Chapter_12#One-point_calibration One-point calibration.

Solution[edit]

Re-open your Radiated Immunity calibration test file and go to 'Field Probes' and make sure that at least one of the probes has the averaging checkbox checked.