Radimation is searching for field sensor but no sensor in the test site
RadiMation is searching for field sensor but no sensor in the test site, why is this happening?
How to fix the problem
In the configuration window of the test goto:
Deselect all the field sensors in this window.
If the check boxes are greyed out, close all windows and add the needed field sensor to the website. If you do not have any add the virtual field sensor to the test equipment.
How does it work?
In the test configuration you can specify that you want the field sensor to be measured, even if this is not the device RadiMation is levelling on.
For example:
- Do a calibration on a room.
- Put the EUT in the room
- You put the Field Sensor on top of the EUT
In this example you are measuring correctly, but you just want know how much field you are generating now that the EUT is in the room.
You do this by telling RadiMation in the menu
It then know that you want the use this equipment.