IEC 61000-4-19

From RadiWiki
Revision as of 07:55, 10 September 2014 by Joro (talk | contribs) (Created page with 'The EN/IEC 61000-4-19 standard describes how Conducted Immunity measurements are performed on A.C. power ports in the frequency range from 2 kHz to 150 kHz. The title…')
(diff) ← Older revision | Latest revision (diff) | Newer revision → (diff)
Jump to: navigation, search

The EN/IEC 61000-4-19 standard describes how Conducted Immunity measurements are performed on A.C. power ports in the frequency range from 2 kHz to 150 kHz. The title of this standard is also: "IEC 61000-4-19: Electromagnetic Compatibility (EMC) - Part 4-19: Testing and measurement techniques - Test for immunity to conducted, differential mode disturbances and signalling in the frequency range from 2 kHz to 150 kHz, at a.c. power ports"

Relevant products:

  • All equipment operating at a mains supply voltage up to 280 V and a frequency of 50 Hz or 60 Hz.

The 'Van der Hoofden' measurement should be performed using a specifically described test-head. This can for example be the Schwarzbeck VDHH 9502 or the AFJ VDH 30. Then the emission measurement should be performed using the following parameters:

Frequency Range 6 dB RBW bandwidth Measurement time Frequency step Detector
20 kHz - 150 kHz 200 Hz 100 ms 220 Hz Peak
150 kHz - 10 MHz 9 kHz 20 ms 10 kHz Peak

The result of the emission measurement results should be used in a calculation to determine the Factor 'F'. The final value of this Factor 'F' should not exceed 0.85 for the measurements to pass.

Related standards

Links