= Measurement settings =
+
#REDIRECT [[Chapter_14#7._Measurement_Settings]]
 
 
With measurement settings you can change the behaviour of the at certain points. Be very careful what you do with these setting.
 
 
 
All settings done here are globally used inside RadiMation.
 
 
 
[[Image:Measurement Settings.PNG]]
 
 
 
 
 
== Starting level signal generator ==
 
With the starting level signal generator you can define the level on which RadiMation is setting the signal generator on at the start of a immunity test.
 
 
 
== Perform power delivery test after pause ==
 
Perform power delivery test after pause is very helpful is you are changing the RF path during the pause.
 
 
 
some examples:
 
*Changing the antenna, it will show high refelected power if the antenna is not correctly connected
 
*Changing the switch matrix to ensure a safe working area while changing something on the [[EUT]]. It will show you if you are injecting power inside the chamber
 
 
 
== Disable frequency zoom-in in emission test ==
 
Disable frequency zoom-in in emission test can be used to decrees the measuring time by not trying to pin point the most accurate frequency. Some labs are only interested in the overall envelope of the emission, for these labs this '''can''' be a very useful option .
 
 
 
{{Warning|Disabling the zoom-in in emmission test in combination with analysers is very likely to result in wrong measurement. }}
 
 
 
== Calibration ==
 
When using the calibration date for devices, you can specify the behaviour here. This may vary between a simple message untill stopping measurements being done.
 
 
 
= Required environment settings =
 
With the required environment settings you can force the test engineer to fill in specific [[Environmental Data]] settings. The first test done with a new EUT document requires the engineer to fill in the requested information. The second time a button will appear that will fill in the same information as the previous time.
 
 
 
 
 
 
 