Difference between revisions of "RadiMation Application Note 123"

From RadiWiki
Jump to: navigation, search
(How to use the switch matrix)
Line 75: Line 75:
  
 
[[File:SwitchMatrix.png]]
 
[[File:SwitchMatrix.png]]
 +
 +
We will show one example in detail. For the other switches, the process is the same.
 +
For explaining purposes I will use the switch design shown in the schematic drawing.
 +
 +
If we look at the first device in the switch design we see that we need to be able to switch between the two amps available in the system. We want to use the amplifier suitable for low frequency for the tests from 80MHz - 1 GHz (AMP1). For the range 1 GHz to 3 GHz we need to switch to the amp for the higher range 1 GHz-3Ghz (AMP2).
 +
 +
Looking at our switch-design we see that AMP1 is Connected to the normally closed output of the switch. Amp2 is connected to the normally open output of the switch. No we have the data we need to set the drivers up correctly.
 +
 +
First make two drivers for the switch settings for AMP1 and AMP2.
 +
 +
Go to device drivers and click "add"

Revision as of 12:14, 12 June 2019

RadiMation Application Note 123


Implementing switch matrix design in RadiMation®

This application note describes how to implement a switch matrix design into RadiMation®

When making a test setup it can be needed to switch to different equipment for different settings or frequency ranges. This is supported in RadiMation® by being able to make a specific driver for a switch matrix to be used in a specific test-site. Later in this text this will hopefully become more clear.

A switch matrix design is normally a drawing or something like a excel sheet. It will tell you what is connected when the relays are in a certain state. The relays normally can be in two modes. Normally open and Normally closed.

Open means that the switch is off and closed means that the switch is on. When we have the switch matrix design we know the signal flow and which switches need to be set to create the desired signal path.


How to use the switch matrix

To use a switch matrix you need to add them with the needed settings to the test site.

Below we will explain an example use-case scenario. We have chosen to show how to implement a specific scenario instead of explaining all the different steps in the process. This is because it is very case specific how you would set it up.


By explaining a scenario you get an idea why it is set-up in a specific way

Scenario:

You have a RI test that you need to set up. The path and devices needed in the different bands need to be different, also we want to use the same measurement equipment for both paths. This means that you need to make two test sites.

One for the low range and one for the high range. In this way you can select the right test-site for the frequency range you are working on. If the test-site is properly set up it will automatically set all the relays need to create the path requested.


Let us further specify the scenario:


We have two ranges with the following equipment:


80 MHz-1 GHz

AMP 200w 80 MHz - 1 GHz

Coupler 80 MHz - 1 GHz

Antenna 80 Mhz- 1 Ghz

1 GHz - 3 GHz

AMP 200w 80 MHz - 1 GHz

Coupler 80 MHz - 1 GHz

Antenna 80 Mhz- 1 Ghz

Apart from this equipment we also have two power meters to measure our forward and reflected power. Now that we know all the parameters we can start to specify our design.


We need two signal paths but want be able to measure both with the same power meter.


See the picture below for a schematic overview how this could be set up.

Switches.png


Now we have our design we can start to translate the design back to Radimation and start to configure the different components. In this specific case we will use separate switches. It is also possible to use switchboards that have more then one relay.


Configuring Radimation to use the right switching when needed

Now that the design is thought out we can configure Radimation.

First start with adding the switches and configure them correctly. Below you will see a picture of the device driver switch matrix,

SwitchMatrix.png

We will show one example in detail. For the other switches, the process is the same. For explaining purposes I will use the switch design shown in the schematic drawing.

If we look at the first device in the switch design we see that we need to be able to switch between the two amps available in the system. We want to use the amplifier suitable for low frequency for the tests from 80MHz - 1 GHz (AMP1). For the range 1 GHz to 3 GHz we need to switch to the amp for the higher range 1 GHz-3Ghz (AMP2).

Looking at our switch-design we see that AMP1 is Connected to the normally closed output of the switch. Amp2 is connected to the normally open output of the switch. No we have the data we need to set the drivers up correctly.

First make two drivers for the switch settings for AMP1 and AMP2.

Go to device drivers and click "add"