= How to program the RadiSense 10 Calibration table =
 
= How to program the RadiSense 10 Calibration table =
In this application note we explain how you can program the RadiSense 10 Calibration table in RadiMation.
+
This {{RadiMation}} application note explains how {{RadiMation}} can be used to program the calibration data in a RadiSense 10 field probe.
  
 
== Programming the table  ==
 
== Programming the table  ==
 +
# Create a RadiMation correction file, by selecting from the menu: {{Menu|File|New|Correction}}.<BR /><BR />[[File:CreateCorrection.png]]
 +
# Add the {{ScreenElement|Frequency}}, {{ScreenElement|X-axis correction}}, {{ScreenElement|Y-axis correction}} and {{ScreenElement|Z-axis correction}} columns.<BR />[[File:AddColumns.png]]
 +
# Paste the earlier determined correction values from Excel into the correction file.<BR />Note. Multiple columns can be pasted from excel into a correction file with a single operation.<BR />Note. If the first row in the table seen in the correction file is not overwritten, remove this row.<BR />[[File:ResultingCorrectionFile.png]]
 +
# Store the correction file.
 +
# Go to the "advanced settings" of the RadiSense device driver and select {{ScreenElement|User Calibration}}".<BR />[[File:ProgrammingCorrectionTable.png]]
 +
# Make sure to check the checkbox “Use calibration if available” if the calibration should be used.
 +
# Click {{ScreenElement|Browse}}. Then find and select the earlier created correction file. Then click {{ScreenElement|Program calibration table}}.
 +
# When the progress bar is at 100% the programming has been succesfully completed.
  
'''1) Create a RadiMation correction file.</BR>'''
 
[[File:CreateCorrection.png]]</BR></BR>
 
'''2) Add the frequency, X-axis-correction, Y-axis-correction and Z-axis-correction columns.</BR>'''
 
[[File:AddColumns.png]]</BR></BR>
 
'''3) Paste the earlier determined correction values from exel into the correction file.</BR>'''
 
Note. Multiple columns can be pasted from excel into a correction file with a single operation.</BR>
 
Note. If the first row in the table seen in the correction file is not overwritten, remove this row.</BR>'''
 
[[File:ResultingCorrectionFile.png]]</BR></BR>
 
'''4) Store the correction file.</BR>'''
 
'''5)  Go to the "advanced settings" of the RadiSense device driver and select "user calibration".</BR>'''
 
[[File:ProgrammingCorrectionTable.png]]</BR></BR>
 
'''6) Make sure to check the checkbox “Use calibration if available” if the calibration should be used.</BR>'''
 
'''7) Click browse. Then find and select the earlier created correction file.</BR>'''
 
'''8) Click “Program calibration table”.</BR>'''
 
'''9) When the progress bar is at 100% the programming has been succesfully completed.</BR>'''
 
  
 +
== Conclusion ==
 +
By following the steps that are explained in this {{RadiMation}} application note, it is possible to program the calibration data into a RadiSense 10 field probe. This has the advantage that all the individual axis are corrected with their own calibration data in the probe itself. Also the isotropic field value is then optimally corrected, and will result in a more accurate measurement.
  
 
[[Category:RadiMation Application Note]]
 
[[Category:RadiMation Application Note]]
 
[[Category:RadiMation]]
 
[[Category:RadiMation]]

Revision as of 13:30, 8 November 2023

How to program the RadiSense 10 Calibration table

This RadiMation® application note explains how RadiMation® can be used to program the calibration data in a RadiSense 10 field probe.

Programming the table

  1. Create a RadiMation correction file, by selecting from the menu:
   Menu.png File
      Menu.png New
         Menu.png Correction .

CreateCorrection.png
  1. Add the Frequency, X-axis correction, Y-axis correction and Z-axis correction columns.
    AddColumns.png
  2. Paste the earlier determined correction values from Excel into the correction file.
    Note. Multiple columns can be pasted from excel into a correction file with a single operation.
    Note. If the first row in the table seen in the correction file is not overwritten, remove this row.
    ResultingCorrectionFile.png
  3. Store the correction file.
  4. Go to the "advanced settings" of the RadiSense device driver and select User Calibration".
    ProgrammingCorrectionTable.png
  5. Make sure to check the checkbox “Use calibration if available” if the calibration should be used.
  6. Click Browse. Then find and select the earlier created correction file. Then click Program calibration table.
  7. When the progress bar is at 100% the programming has been succesfully completed.


Conclusion

By following the steps that are explained in this RadiMation® application note, it is possible to program the calibration data into a RadiSense 10 field probe. This has the advantage that all the individual axis are corrected with their own calibration data in the probe itself. Also the isotropic field value is then optimally corrected, and will result in a more accurate measurement.