RadiMation Error: 536871941

From RadiWiki
Revision as of 15:50, 31 October 2016 by Joro (talk | contribs) (Created page with '= Start Frequency is too low for the selected equipment. Usable frequency range of '<device>' is from <start-frequency> until <stop-frequency> During the initialization of the te…')
(diff) ← Older revision | Latest revision (diff) | Newer revision → (diff)
Jump to: navigation, search

= Start Frequency is too low for the selected equipment. Usable frequency range of '<device>' is from <start-frequency> until <stop-frequency> During the initialization of the testsite, the allowed frequency range of the used equipment ("<device>") is checked. This error is raised if it is detected that the device is limited in the frequency range, and cannot be used for the complete frequency range of the test. Because of this problem the test cannot be started.

Solution

  • Reduce the frequency range of the test. For example by increasing the start frequency of the test, or decreasing the stop frequency of the test
  • Extend the allowed frequency range of the device.
  • Check the minimum and maximum values in the frequency column of the correction files attached to the device.