Revision as of 12:34, 6 June 2019

There is no average probe selected

This error happens when you are doing a Radiated Immunity calibration and do not have at least one averaging probe selected. For a RI calibration at least one averaging probe is mandatory to do the calibration.

Solution

Re-open your Radiated Immunity calibration test file and go to 'Field Probes' and make sure that at least one of the probes has the averaging checkbox checked.