RadiMation Error: 536871966

From RadiWiki
Revision as of 12:33, 6 June 2019 by Tjsc (talk | contribs) (Created page with "= There is no average probe selected = This error happens when you are doing a Radiated Immunity calibration and do not have at least one averaging probe selected. For a RI...")
(diff) ← Older revision | Latest revision (diff) | Newer revision → (diff)
Jump to: navigation, search

There is no average probe selected

This error happens when you are doing a Radiated Immunity calibration and do not have at least one averaging probe selected. For a RI calibration at least one averaging probe is mandatory to do the calibration.

Solution

Re-open your Radiated Immunity calibration test file and go to ('<Field Probe>')and make sure that at least one of the probes has the averaging checkbox checked.