Difference between revisions of "RadiMation Error: 536871984"

From RadiWiki
Jump to: navigation, search
(Expert user solutions)
(Expert user solutions)
Line 15: Line 15:
 
In case you are an advanced user and know what you are doing, it is also possible to adjust advanced settings of this feature.
 
In case you are an advanced user and know what you are doing, it is also possible to adjust advanced settings of this feature.
  
To do so open the TUNING.INI in the CONFDVDR folder and search for the table [PowerDelivery]. Be careful as this concerns a safety feature.  
+
To do so open the TUNING.INI in the CONFDVDR folder and search for the table [PowerDelivery]. Be very careful as this concerns a safety feature.  
 
* Set the 'CheckReflectedPower' to False.
 
* Set the 'CheckReflectedPower' to False.
 
* Decrease the 'NeededForwardReflectedPowerDifference' to a different value in dB.
 
* Decrease the 'NeededForwardReflectedPowerDifference' to a different value in dB.

Revision as of 13:01, 13 October 2016

Too Much Reflected Power

The error Too Much Reflected Power is raised after the power delivery test has failed. The differences between the Forward and Reflected powermeter values is determined by measuring the Forward and Reflected power meters during the Test Initialization. RadiMation® requires that the Forward powermeter value is at least >= 6.0 dB then the measured Reflected powermeter value. If this condition is not met it will cause this error.

Solutions

  • Inspect the current test setup if there are no unusual malfunctions.
  • Ensure that no bad or broken cables are used.
  • Ensure that (if applicable) corrections are applied for both Forward and Reflected powermeter in RadiMation.
  • Remove the Reflected Powermeter from the used test equipment (and eventually disable the Amplifier protection Reflected powermeter from the TSF).
  • Manually set the Signal generator to the correct frequency (normally the center of the configured test frequency range) and carrier level and measure both Forward as Reflected power. Is the difference between these manual measurements also < 6 dBm?

Expert user solutions

In case you are an advanced user and know what you are doing, it is also possible to adjust advanced settings of this feature.

To do so open the TUNING.INI in the CONFDVDR folder and search for the table [PowerDelivery]. Be very careful as this concerns a safety feature.

  • Set the 'CheckReflectedPower' to False.
  • Decrease the 'NeededForwardReflectedPowerDifference' to a different value in dB.
  • To completely disable this functionality (not recommended) set 'Enabled' to False.