RadiMation Error: 536879952

From RadiWiki
Revision as of 10:07, 3 August 2007 by Joro (talk | contribs)
Jump to: navigation, search

The device driver system is currently not usable

RadiMation has detected that another RadiMation instance is also using one or more device drivers to communicate with equipment. This situation is detected during the startup of RadiMation, and probably the error RadiMation Error: 805306782 is also shown. To prevent that the two (or more) simultaneously running programs are changing equipment settings of each other it is not allowed to use the device driver system from the second (or more) running RadiMation process.

Solution

  • Switch to the other running RadiMation program, and perform the test in the other instance of RadiMation
  • Only start one RadiMation program at the same time.
  • Close all RadiMation programs that are available in the taskbar and start RadiMation only once.
  • Terminate or Kill all 'RADIMAT-MAIN.EXE' processes in the task manager and start RadiMation only once.
  • Restart the PC and then start RadiMation only once.

If 2 RadiMation instances are started, closing the first one does not give the second instance control over the GPIB board. This is correct functionality. During the start up of RadiMation there is a check to see if the board is available. If this check is unsuccessful, it will be remembered as long as RadiMation is running. Close all RadiMation programs to make the device driver system usable again.