Difference between revisions of "RadiMation Error: 536879959"

From RadiWiki
Jump to: navigation, search
(Created page with '= Could not create a LiteX database connection. Is the LiteX library installed? = {{RadiMation}} is using a LiteX database for faster searching through lists. It seems that no Li…')
 
(Could not create a LiteX database connection. Is the LiteX library installed?)
 
Line 1: Line 1:
= Could not create a LiteX database connection. Is the LiteX library installed? =
+
= Could not create a SQLite database connection. Is the SQLite library installed? =
{{RadiMation}} is using a LiteX database for faster searching through lists. It seems that no LiteX database could be created.  
+
{{RadiMation}} is using a SQLite database for faster searching through lists. It seems that no SQLite database could be created.
 +
 
 +
It also can happen that this error is caused by a conflicting DLL that is loaded by the Keysight IO Libraries suite. Older versions of the Keysight IO Libraries suite that are used to control a Keysight GPIB card to control the measurement equipment, can also cause this error. More recent versions (>= 18.3.29324.3) of the Keysight IO Libraries suite do not trigger this error anymore.
  
 
== Solution ==
 
== Solution ==
 
* Reinstall the {{RadiMation}} software.
 
* Reinstall the {{RadiMation}} software.
* Check if the file SQLITE3.DLL is available on your PC.
+
* If you have the Keysight IO Libraries suite installed: Update to version 18.3.29324.3 or newer
  
 
[[Category:RadiMation]]
 
[[Category:RadiMation]]
 
[[Category:RadiMation Error]]
 
[[Category:RadiMation Error]]

Latest revision as of 10:07, 7 July 2023

Could not create a SQLite database connection. Is the SQLite library installed?

RadiMation® is using a SQLite database for faster searching through lists. It seems that no SQLite database could be created.

It also can happen that this error is caused by a conflicting DLL that is loaded by the Keysight IO Libraries suite. Older versions of the Keysight IO Libraries suite that are used to control a Keysight GPIB card to control the measurement equipment, can also cause this error. More recent versions (>= 18.3.29324.3) of the Keysight IO Libraries suite do not trigger this error anymore.

Solution

  • Reinstall the RadiMation® software.
  • If you have the Keysight IO Libraries suite installed: Update to version 18.3.29324.3 or newer