− | = Cannot find spectrum | + | = Cannot find spectrum analyser in the currently active testsite = |
− | As part of the test, an operation on the spectrum | + | As part of the test, an operation on the spectrum analyser is needed. However the currently selected test-site doesn't have a device selected for the spectrumn analyser. |
− | It could also mean that during the initialization of the spectrum | + | It could also mean that during the initialization of the spectrum analyser, an error occurred, and that no active spectrum analyser is present. |
== Solution == | == Solution == | ||
− | * Check the configuration of the used test-site to ensure that it includes a spectrum | + | * Check the configuration of the used test-site to ensure that it includes a spectrum analyser |
− | * Ensure that no errors happened during the initialization of the spectrum | + | * Ensure that no errors happened during the initialization of the spectrum analyser |
[[Category:RadiMation]] | [[Category:RadiMation]] | ||
[[Category:RadiMation Error]] | [[Category:RadiMation Error]] |
Latest revision as of 14:49, 22 September 2022
Cannot find spectrum analyser in the currently active testsite
As part of the test, an operation on the spectrum analyser is needed. However the currently selected test-site doesn't have a device selected for the spectrumn analyser.
It could also mean that during the initialization of the spectrum analyser, an error occurred, and that no active spectrum analyser is present.
Solution
- Check the configuration of the used test-site to ensure that it includes a spectrum analyser
- Ensure that no errors happened during the initialization of the spectrum analyser