= The used scanning receiver doesn't support a logarithmic step =
 
= The used scanning receiver doesn't support a logarithmic step =
This error is raised when a logarithmic stepsize is configured in the TSF file, but not supported by the used equipment.
+
This error is raised due to hardware limitations in a situation a logarithmic stepsize is configured in the TSF file, but not supported by the used equipment.
  
 
== Solution ==
 
== Solution ==
* Close the message and abort this test. Use restart last TSF to have quick access to modify the stepsize to a different method e.g. Linear or Auto. Press Start Test.
+
* Close the message and abort this test. Use restart last TSF to have quick access to modify the stepsize to a different method for example to <b>Linear</b> or <b>Auto</b>. Press Start Test.
  
 
[[Category:RadiMation Error]]
 
[[Category:RadiMation Error]]
 
[[Category:RadiMation]]
 
[[Category:RadiMation]]

Latest revision as of 10:08, 22 July 2014

The used scanning receiver doesn't support a logarithmic step

This error is raised due to hardware limitations in a situation a logarithmic stepsize is configured in the TSF file, but not supported by the used equipment.

Solution