Difference between revisions of "RadiMation Error: 536881992"
(→Solution) |
(→Solution) |
||
(One intermediate revision by the same user not shown) | |||
Line 1: | Line 1: | ||
= The used scanning receiver doesn't support a logarithmic step = | = The used scanning receiver doesn't support a logarithmic step = | ||
− | This error is raised | + | This error is raised due to hardware limitations in a situation a logarithmic stepsize is configured in the TSF file, but not supported by the used equipment. |
== Solution == | == Solution == | ||
− | * Close the message and abort this test. Use restart last TSF to have quick access to modify the stepsize to a different method | + | * Close the message and abort this test. Use restart last TSF to have quick access to modify the stepsize to a different method for example to <b>Linear</b> or <b>Auto</b>. Press Start Test. |
[[Category:RadiMation Error]] | [[Category:RadiMation Error]] | ||
[[Category:RadiMation]] | [[Category:RadiMation]] |
Latest revision as of 10:08, 22 July 2014
The used scanning receiver doesn't support a logarithmic step
This error is raised due to hardware limitations in a situation a logarithmic stepsize is configured in the TSF file, but not supported by the used equipment.
Solution
- Close the message and abort this test. Use restart last TSF to have quick access to modify the stepsize to a different method for example to Linear or Auto. Press Start Test.