Difference between revisions of "RadiMation Error: 805329921"

From RadiWiki
Jump to: navigation, search
(Solution)
(Amplifier returned that the RF loop is not allocated to the currently active RF path and is not evaluated.)
 
(One intermediate revision by the same user not shown)
Line 1: Line 1:
= Amplifier Interlock query returned an unexpected respond. =
+
= Amplifier returned that the Group Interlock [number] is not allocated to the currently active RF path and is not evaluated. =
It seems that the amplifier Interlock query returned an unexpected respond.
+
It seems that the amplifier Interlock query returned that the RF loop is not allocated to the currently active RF path and is not evaluated.
  
 
== Solution ==
 
== Solution ==
* In case this problems occurs at regular bases please report this [[issue]] to your [[reseller]] or to DARE!! Instruments (fill in the [http://www.dare.eu/en/instruments/emc-test-software/radimation-issue-report online form] or mail to {{MailSupport}}) and include a detailed description how to reproduce this issue.
+
* Check that the correct interlock number is configured in the device settings.
 +
* Verify if this interlock group number is also used for the current RF Path. Only this specific interlock is checked by the Amplifier.
  
 
[[Category:RadiMation Error]]
 
[[Category:RadiMation Error]]
 
[[Category:RadiMation]]
 
[[Category:RadiMation]]

Latest revision as of 15:54, 5 January 2015

Amplifier returned that the Group Interlock [number] is not allocated to the currently active RF path and is not evaluated.

It seems that the amplifier Interlock query returned that the RF loop is not allocated to the currently active RF path and is not evaluated.

Solution

  • Check that the correct interlock number is configured in the device settings.
  • Verify if this interlock group number is also used for the current RF Path. Only this specific interlock is checked by the Amplifier.