Difference between revisions of "RadiMation Error: 805335276"
(→The firmware version of the LPS card is to low send the calibration data) |
|||
Line 1: | Line 1: | ||
− | = The firmware version of the LPS card is too low send the calibration data = | + | = The firmware version of the LPS card is too low to send the calibration data = |
If a RadiSense 10 is connected to a multislot RadiCentre, the LPS card FW has to be V2.6.3 or newer to upload the calibration data | If a RadiSense 10 is connected to a multislot RadiCentre, the LPS card FW has to be V2.6.3 or newer to upload the calibration data |
Revision as of 13:32, 30 March 2022
The firmware version of the LPS card is too low to send the calibration data
If a RadiSense 10 is connected to a multislot RadiCentre, the LPS card FW has to be V2.6.3 or newer to upload the calibration data
Solution
- Update the firmware of the LPS card.