Difference between revisions of "RadiMation Error: 805339670"

From RadiWiki
Jump to: navigation, search
(Created page with '= Quasi Peak detector can only be activated when an EMI RBW bandwidth is used. = The Siglent SSA 3000-series only can use the Quasi peak detector, when it is the only activated …')
 
 
Line 1: Line 1:
 
= Quasi Peak detector can only be activated when an EMI RBW bandwidth is used. =
 
= Quasi Peak detector can only be activated when an EMI RBW bandwidth is used. =
  
The Siglent SSA 3000-series only can use the Quasi peak detector, when it is the only activated detector. We detected that multiple detectors are activated, of which one is the Quasi Peak detector.
+
The Siglent SSA 3000-series only can use the Quasi peak detector, when the EMI RBW filter has been selected. This error is raised when the Gaussian RBW filter is selected.
  
 
== Solution ==
 
== Solution ==
* Change the configuration of the to be measured traces, to only have the Quasi Peak detector activated.
+
* Change the configuration of test to use a RBW of 200 Hz, 9 kHz or 120 kHz. The driver will automatically select the EMI RBW filter.
  
 
[[Category:RadiMation]]
 
[[Category:RadiMation]]
 
[[Category:RadiMation Error]]
 
[[Category:RadiMation Error]]

Latest revision as of 12:22, 16 November 2017

Quasi Peak detector can only be activated when an EMI RBW bandwidth is used.

The Siglent SSA 3000-series only can use the Quasi peak detector, when the EMI RBW filter has been selected. This error is raised when the Gaussian RBW filter is selected.

Solution

  • Change the configuration of test to use a RBW of 200 Hz, 9 kHz or 120 kHz. The driver will automatically select the EMI RBW filter.