Difference between revisions of "RadiMation Error: 805351271"

From RadiWiki
Jump to: navigation, search
(Created page with "= The selected RBW, is not possible when the EMC Option is not available, or when the tracking generator option is enabled = When the EMC option is not available. The accepted...")
 
 
Line 1: Line 1:
 
= The selected RBW, is not possible when the EMC Option is not available, or when the tracking generator option is enabled =
 
= The selected RBW, is not possible when the EMC Option is not available, or when the tracking generator option is enabled =
When the EMC option is not available. The accepted values for RBW is limited. This is also the case when tracking generator is used.
+
When the EMC option is not available, the CISPR filter bandwidths for the RBW are not available in the analyser.
 +
 
 +
Also when the tracking generator option is activated, the CISPR filter bandwidths cannot be activated.
  
 
== Solution ==
 
== Solution ==
 
* Try changing the selected RBW value's in the test configuration.
 
* Try changing the selected RBW value's in the test configuration.
 +
* Turn off the tracking generation option for the analyser.
  
 
[[Category:RadiMation Error]]
 
[[Category:RadiMation Error]]
 
[[Category:RadiMation]]
 
[[Category:RadiMation]]

Latest revision as of 11:34, 21 June 2019

The selected RBW, is not possible when the EMC Option is not available, or when the tracking generator option is enabled

When the EMC option is not available, the CISPR filter bandwidths for the RBW are not available in the analyser.

Also when the tracking generator option is activated, the CISPR filter bandwidths cannot be activated.

Solution

  • Try changing the selected RBW value's in the test configuration.
  • Turn off the tracking generation option for the analyser.