Difference between revisions of "RadiMation Error: 805366369"

From RadiWiki
Jump to: navigation, search
(Created page with "Because the RBW values have a different values in these bands the test needs to be split the test into multiple bands: Band A: 9 kHz - 150 kHz (200 Hz) Band B: 150 kHz - 30 M...")
 
 
(2 intermediate revisions by the same user not shown)
Line 1: Line 1:
Because the RBW values have a different values in these bands the test needs to be split the test into multiple bands:
+
==Frequency span with multiple RBW values setup==
  
Band A: 9 kHz - 150 kHz (200 Hz)
+
==Solution==
Band B: 150 kHz - 30 MHz (9 kHz)
+
Because the RBW values have a different values in these bands the test needs to be split the test into multiple bands or separate tests:
Band C: 30 MHz - 1 GHz (120 kHz)
+
 
 +
* Band A: 9 kHz - 150 kHz (200 Hz)
 +
* Band B: 150 kHz - 30 MHz (9 kHz)
 +
* Band C: 30 MHz - 1 GHz (120 kHz)
 +
 
 +
[[Category:RadiMation]]
 +
[[Category:RadiMation Error]]

Latest revision as of 13:03, 13 April 2022

Frequency span with multiple RBW values setup

Solution

Because the RBW values have a different values in these bands the test needs to be split the test into multiple bands or separate tests:

  • Band A: 9 kHz - 150 kHz (200 Hz)
  • Band B: 150 kHz - 30 MHz (9 kHz)
  • Band C: 30 MHz - 1 GHz (120 kHz)