Difference between revisions of "RadiMation Error: 805366369"
(Created page with "Because the RBW values have a different values in these bands the test needs to be split the test into multiple bands: Band A: 9 kHz - 150 kHz (200 Hz) Band B: 150 kHz - 30 M...") |
|||
(2 intermediate revisions by the same user not shown) | |||
Line 1: | Line 1: | ||
− | + | ==Frequency span with multiple RBW values setup== | |
− | Band A: 9 kHz - 150 kHz (200 Hz) | + | ==Solution== |
− | Band B: 150 kHz - 30 MHz (9 kHz) | + | Because the RBW values have a different values in these bands the test needs to be split the test into multiple bands or separate tests: |
− | Band C: 30 MHz - 1 GHz (120 kHz) | + | |
+ | * Band A: 9 kHz - 150 kHz (200 Hz) | ||
+ | * Band B: 150 kHz - 30 MHz (9 kHz) | ||
+ | * Band C: 30 MHz - 1 GHz (120 kHz) | ||
+ | |||
+ | [[Category:RadiMation]] | ||
+ | [[Category:RadiMation Error]] |
Latest revision as of 13:03, 13 April 2022
Frequency span with multiple RBW values setup
Solution
Because the RBW values have a different values in these bands the test needs to be split the test into multiple bands or separate tests:
- Band A: 9 kHz - 150 kHz (200 Hz)
- Band B: 150 kHz - 30 MHz (9 kHz)
- Band C: 30 MHz - 1 GHz (120 kHz)