RadiMation is searching for field sensor but no sensor in the test site, why is this happening?
+
#REDIRECT [[FAQ#Why is an immunity test trying to control a field sensor, if no field sensor is selected in the used testsite?]]
 
 
=How to fix the problem=
 
 
 
In the configuration window of the test goto:
 
{{menu|Inputs|Field probes}}
 
Deselect all the field sensors in this window.
 
 
 
If the check boxes are greyed out, close all windows and add the needed [[field sensor]] to the website.
 
If you do not have any add the [[virtual field sensor]] to the [[test equipment]].
 
 
 
=How does it work?=
 
 
 
In the test configuration you can specify that you want the field sensor to be measured, even if this is not the device RadiMation is levelling on.
 
 
 
For example:
 
*Do a calibration on a room.
 
*Put the [[EUT]] in the room
 
*You put the [[Field Sensor]] on top of the [[EUT]]
 
 
 
In this example you are measuring correctly, but you just want know how much field you are generating now that the [[EUT]] is in the room.
 
 
 
You do this by telling RadiMation in the menu {{menu|Inputs|Field probes}}
 
 
 
It then know that you want the use this equipment.
 
 
 
=Why does it remove it when not configured?=
 
 
 
 
 