Schwarzbeck HFRA 5158

From RadiWiki
Revision as of 12:41, 21 June 2007 by Joro (talk | contribs) (Updated DeviceDriverInfo template)
(diff) ← Older revision | Latest revision (diff) | Newer revision → (diff)
Jump to: navigation, search

The Schwarzbeck HFRA 5158 device driver is a Antenna which is supported by RadiMation®.

Configuration

The following tabs are available in the advanced configuration of the Schwarzbeck HFRA 5158:

ConfigurableFrequencyRangeTab.png

The frequency range of the Schwarzbeck HFRA 5158 as provided by the manufacturer is shown and selected as default. It is possible to overrule these frequencies and to manual adjust the allowed frequency range of the Schwarzbeck HFRA 5158.

ScreenElementDescription.png Use default start frequency If the checkbox is checked, the default start frequency will be used as the lowest usable frequency in a test for this device.
ScreenElementDescription.png Custom start frequency If the Use default start frequency checkbox is unchecked, another start frequency (expressed in MHz) can be specified. The customized start frequency will then be used as the lowest usable frequency in a test for this device. The customized frequency can be a limitation or an extension of the default start frequency.
ScreenElementDescription.png Use default stop frequency If the checkbox is checked, the default stop frequency will be used as the highest usable frequency in a test for this device.
ScreenElementDescription.png Custom stop frequency If the Use default stop frequency checkbox is unchecked, another stop frequency (expressed in MHz) can be specified. The customized stop frequency will then be used as the highest usable frequency in a test for this device. The customized frequency can be a limitation or an extension of the default stop frequency.


Specifying a different frequency range can be useful if for example:

  • A device (like a coupler, antenna, injection device, cable, etc...) is still useable (but out of specification) outside the standard suggested frequency range.
  • An external mixer is used to measure an extended frequency range.
  • An up- or down-convertor is used to shift the frequency range.
  • A newer model of a device is present that has an extended frequency range, and still uses the same remote control commands.

Be careful changing these setting as RadiMation® is no longer able to verify if the Schwarzbeck HFRA 5158 is used outside frequency range that is specified by the manufacturer. This may result to serious damage of your measurement device.


AntennaSettings.png

Information.png
Note: The antenna efficiency factor is only applicable to reverberation chamber testing.


ScreenElementDescription.png Efficiency The antenna efficiency factor of the antenna.

The antenna efficiency factor is assumed (if not known) to be 0.75 for a log periodic antenna and 0.9 for a horn antenna. It is a unitless factor with values between 1.00 and 0.00. The antenna efficiency factor is a single fixed value which is used for all the frequencies of the antenna. Values smaller than 0.00 or larger than 1.00 are not allowed and the configuration dialog of the antenna cannot be closed if an invalid value is specified.

An antenna efficiency correction factor can also be selected in the device driver settings dialog in RadiMation. That correction file contains a value that is a correction that is added on top of the value from the device driver. Negative values down to -1.0 are also allowed as it is a correction on the single fixed value from the antenna driver, and it allows to correct that single fixed value to a lower value. The combination of the single fixed antenna efficiency value from the antenna driver, and the added correction value from the selected correction file, should be between 1.0 and 0.0. If this combined value is invalid for one or more frequencies a specific error is shown during the initialisation of the test.