| U_L_N 325 | | U_L_N 325 |
Revision as of 08:13, 6 October 2016
The Configurable AD Convertor device driver is a AD Converter which is supported by RadiMation®.
Reset | The reset code that needs to be send to device. When left blank no command will be send. |
Init | The Init code that needs to be send to device. When left blank no command will be send. |
Get Id | The code that needs to be send to device to get the identification back. A common used SCPI command is *IDN?. When left blank no command will be send. |
Trigger and Reading |
|
Trigger and Reading |
|
Readback example(s)[edit]
Received information
|
Regex
|
RadiMation Readout
|
Note
|
---|---|---|---|
U_L_N 325 | (-?[0-9.,Ee-]+)
|
325
|
Takes the first number. |
I_L1;12.34 | ;(-?[0-9.,Ee-]+)
|
12.34
|
Takes the first number after ";". |
THD_U_L1;;14,5 | ;;(-?[0-9.,Ee-]+)
|
14.5
|
Takes the first number after ";;". |
AC_FREQ;Channel1;1.23E3 | ;*;(-?[0-9.,Ee-]+)
|
1230
|
Takes the first number after the appearance of a second ";". |