RadiMation Application Note 133: Difference between revisions

From RadiWiki
Jump to navigation Jump to search
(Created page with "= How to use EUT monitoring in Pulsed Immunity MultiBand EFT & Surge {{RadiMation}} = In this application note we shall explain how {{RadiMation}} can be configured to measure...")
 
 
(2 intermediate revisions by the same user not shown)
Line 1: Line 1:
= How to use EUT monitoring in Pulsed Immunity MultiBand EFT & Surge {{RadiMation}} =
= How to use EUT monitoring in Pulsed Immunity MultiBand EFT & Surge {{RadiMation}} =
In this application note we shall explain how {{RadiMation}} can be configured to measure EUT monitoring inputs in Pulsed Immunity MultiBand EFT & Surge.
This application note describes how {{RadiMation}} can be configured to measure EUT monitoring inputs in Pulsed Immunity MultiBand EFT & Surge.


== EUT monitoring in Pulsed Immunity MultiBand EFT & Surge ==
== EUT monitoring in Pulsed Immunity MultiBand EFT & Surge ==
The EFT/Surge MB tests have the abillity to measure EUT monitoring inputs during the active test sequences.</br>
The EFT/Surge MB tests have the abillity to measure EUT monitoring inputs during the active test sequences.</br>
The measured data can be seen in a graph and table view on the Data-tab.</br>
The measured data can be seen in a graph and table view on the Data-tab.</br>
If the min/max range is enabled for the monitor inputs, RadiMation shall display a list of the measurements measured out of range on the Peaks-tab found on the Data window.</br>
If the min/max range is enabled for the monitor inputs, RadiMation will display a list of the measurements measured out of range on the Peaks-tab found on the Data window.</br>
{{ReportGeneratorInfo|For more information on how to configurate the EUT monitor inputs, see https://wiki.radimation.com/wiki/index.php/Chapter_10}}
{{ReportGeneratorInfo|For more information on how to configurate the EUT monitor inputs, see https://wiki.radimation.com/wiki/index.php/Chapter_10}}
[[File:Pulsed Imm EUTMonitoring.JPG]]
[[File:Pulsed Imm EUTMonitoring.JPG]]
Line 12: Line 12:


To indicate the stable situation during an EFT/SURGE MB tests an ambient measurement phase is performed before the start of each configured sequence.</br>
To indicate the stable situation during an EFT/SURGE MB tests an ambient measurement phase is performed before the start of each configured sequence.</br>
The ambient measurement phase shall be performed over the default duration of '''3''' seconds.</br>
The ambient measurement phase is performed over the default duration of '''3''' seconds.</br>


[[Category:RadiMation Application Note]]
[[Category:RadiMation Application Note]]
[[Category:RadiMation]]
[[Category:RadiMation]]

Latest revision as of 11:44, 2 July 2021

How to use EUT monitoring in Pulsed Immunity MultiBand EFT & Surge RadiMation®[edit]

This application note describes how RadiMation® can be configured to measure EUT monitoring inputs in Pulsed Immunity MultiBand EFT & Surge.

EUT monitoring in Pulsed Immunity MultiBand EFT & Surge[edit]

The EFT/Surge MB tests have the abillity to measure EUT monitoring inputs during the active test sequences.
The measured data can be seen in a graph and table view on the Data-tab.
If the min/max range is enabled for the monitor inputs, RadiMation will display a list of the measurements measured out of range on the Peaks-tab found on the Data window.

Report generator.png
For more information on how to configurate the EUT monitor inputs, see https://wiki.radimation.com/wiki/index.php/Chapter_10

Pulsed Imm EUTMonitoring.JPG

Ambient Measurement[edit]

To indicate the stable situation during an EFT/SURGE MB tests an ambient measurement phase is performed before the start of each configured sequence.
The ambient measurement phase is performed over the default duration of 3 seconds.