RadiMation Error: 805317370: Difference between revisions
Jump to navigation
Jump to search
No edit summary |
No edit summary |
||
Line 6: | Line 6: | ||
== Solution == | == Solution == | ||
* Please report this [[issue]] to your [[reseller]] or to [[Raditeq]] (fill in the [ | * Please report this [[issue]] to your [[reseller]] or to [[Raditeq]] (fill in the [[Error Report]] or mail to {{MailSupport}}) and include a detailed description how to reproduce this issue. | ||
[[Category:RadiMation Error]] | [[Category:RadiMation Error]] | ||
[[Category:RadiMation]] | [[Category:RadiMation]] |
Latest revision as of 18:47, 29 September 2024
Set frequency ('<x> MHz', '<y> Hz') (diff: '<z> Hz') was not correctly set, Powermeter returned: '<response>'
The R&S SMB is used to measure the power at a specific frequency (at frequency: <x> MHz), however the powermeter is performing the measurement at another frequency (measured frequency: <y> Hz). This results in a measurement frequency error (<z> Hz) that is bigger than is desired. This problem can be caused by an unexpected or erroneous response of the powermeter, which is also shown.
Current frequency was not correctly set
The R&S SMB is used to measure the power at a specific frequency, however the powermeter is performing the measurement at another frequency. The exact frequencies are not known at this moment.
Solution
- Please report this issue to your reseller or to Raditeq (fill in the Error Report or mail to radimation-support@raditeq.com) and include a detailed description how to reproduce this issue.