A very often used verification of the correct functioning of measurement equipment and software, is the measurement of a signal that is generated by a signal generator, using a measurement receiver.

However, in some situations it is possible that a small frequency offset is detected. This page describes a possible solution for the frequency offset when doing a measurement with a signal generator and a measurement receiver.

Assume the following configurations:

Signal Generator Settings
Setting Value
Carrier Frequency 100 kHz
Receiver Settings
Setting Value
Start Frequency 30 kHz
End Frequency 150 kHz
RBW 200 Hz
Step size 200 Hz
Information.png
Note: The selected step frequency of 200 Hz is also possibly generating complications. It is always advised to use the selected RBW/2 or less as the step frequency.

With these settings we get the following emission spectrum: ReceiverOffSetExample.PNG

As you can see the emission is starting a little bit before the 100 kHz. And the measured emission signal is already decreasing on 100 kHz itself.

A possible cause[edit]

A possible cause for this is that the 10 MHz references of both measurement devices are not exactly the same. This can result in small differences in the frequency accuracy.

It is always best to connect the external reference output from one device to the external reference input of the other devices. This will ensure that all devices will use the same 10 MHz reference. It is very often possible for every device, to generate the 10 MHz reference, but in many cases the 10 MHz from the receiver will be more accurate then the one of the signal generator. Because the references are now the same, the above described frequency offset will be more accurate (in the order of Hz). Because the measurement is done with a 200 Hz RBW and step size this may just be the accuracy that is needed to detect the peak at exactly 100 kHz.

Manual[edit]

In the manual of your measurement receiver and signal generator you can find how to configure the devices.

Example connectors[edit]