How to perform a MIL-STD CS114, Conducted susceptibility, bulk cable injection test[edit]
This application note explains how the MIL-STD CS114, Conducted susceptibility, bulk cable injection test can be performed with RadiMation®
The exact requirements and test methods for the CS114 are specified in the MIL-STD-461.
Necessary equipment[edit]
The following devices are necessary to execute this test:
- Signal generator
- Amplifier
- Coupler
- Forward power meter
- Reflected power meter (optionally)
- Current sensor
- Sensor power meter / Analyser
- 50 Ohm load
- 50 Ohm attenuator
- LISN
- Current injection device
- Current injection calibration jig (Calibration fixture)
- Cable drivers with corrections
Configuration of the test sites[edit]
Calibration[edit]
The configuration of the calibration test site should contain the following devices:
Device name |
Tab in testsite configuration window |
note
|
Signal generator |
Devices 1 |
The signal generator to use
|
Amplifier |
Devices 1 |
The amplifier to use
|
Coupler |
Devices 1 |
The coupler to use
|
Forward power meter |
Devices 1 |
The forward power meter to use
|
Reflected power meter |
Devices 1 |
The reflected power meter to use (optionally)
|
Sensor power meter |
Devices 2 |
The power meter or analyser to use for measuring the current
|
Current sensor |
Devices 2 |
The current sensor to use with transfer factor attached to the driver
|
Injection device |
Devices 2 |
The injection clamp to use
|
Jig |
Devices 2 |
The jig to use
|
Cable current -> power meter |
Cables |
Cable (1) with a correction file specified for the cable loss
|
Cable current -> power meter |
Cables |
Cable (2) with the specified loss of the used attenuator
|
Cable current -> power meter |
Cables |
Cable (3) with a correction file specified for the cable loss
|
Verification[edit]
The configuration of the verification test site should contain the following devices:
Device name |
Tab in testsite configuration window |
note
|
Signal generator |
Devices 1 |
The signal generator to use
|
Amplifier |
Devices 1 |
The amplifier to use
|
Coupler |
Devices 1 |
The coupler to use
|
Forward power meter |
Devices 1 |
The forward power meter to use
|
Reflected power meter |
Devices 1 |
The reflected power meter to use (optionally)
|
Sensor power meter |
Devices 2 |
The power meter or analyser to use for measuring the current
|
Current sensor |
Devices 2 |
The current sensor to use with transfer factor attached to the driver
|
Injection device |
Devices 2 |
The injection clamp to use
|
Cable current -> power meter |
Cables |
Cable (1) with a correction file specified for the cable loss
|
Cable current -> power meter |
Cables |
Cable (2) with the specified loss of the used attenuator
|
Cable current -> power meter |
Cables |
Cable (3) with a correction file specified for the cable loss
|
EUT Testing[edit]
The configuration of the eut test site should contain the following devices:
Device name |
Tab in testsite configuration window |
note
|
Signal generator |
Devices 1 |
The signal generator to use
|
Amplifier |
Devices 1 |
The amplifier to use
|
Coupler |
Devices 1 |
The coupler to use
|
Forward power meter |
Devices 1 |
The forward power meter to use
|
Reflected power meter |
Devices 1 |
The reflected power meter to use (optionally)
|
Sensor power meter |
Devices 2 |
The power meter or analyser to use for measuring the current
|
Current sensor |
Devices 2 |
The current sensor to use with transfer factor attached to the driver
|
Injection device |
Devices 2 |
The injection clamp to use
|
Cable current -> power meter |
Cables |
Cable (1) with a correction file specified for the cable loss
|
Cable current -> power meter |
Cables |
Cable (2) with the specified loss of the used attenuator
|
Cable current -> power meter |
Cables |
Cable (3) with a correction file specified for the cable loss
|
Calibration procedure[edit]
Create correction file for the test level[edit]
- File
- New
- Correction
First create a correction file in RadiMation® and specify the applicable test level (See 'TABLE IV CS114 limit curves.' and 'FIGURE CS114-1. CS114 calibration limits.').
In this example the limits for 'Aircraft (External or safety critcal) / Curve 5 are configured.
Make sure to remove the Correction column and add the Current column with the unit set to dBuV.
Create calibration test setup files[edit]
There are three frequency ranges that have to be calibrated with the following settings (see TABLE III).
Band |
Frequency range |
Step size
|
Band 1 |
10 kHz - 1 MHz |
5 %
|
Band 2 |
1 MHz - 30 MHz |
1 %
|
Band 2 |
30 MHz - 200 MHz |
0.5 %
|
- Calibration
- System calibration
- Conducted immunity
Create three test setup files, this example shows the settings for band 1.
Start
|
The start frequency of the test. For example 10 kHz.
|
End
|
The stop frequency of the test. For example 1 MHz.
|
Step
|
The frequency step, in this case 5%.
|
Test level
|
The test level, in this case 5%.
|
Variable
|
The variable test level, select the correction file created earlier.
|
Tolerance
|
The regulation tolerance.
|
Calibration method
|
Forward power.
|
Test equipment
|
Calibration test equipment.
|
Test engineer
|
The engineer that performed the calibration.
|
Verification[edit]
EUT Testing[edit]
Conclusion[edit]