Pre Wait Time
|
Wait time before starting the measurement-loop. Can be used to wait for a signal generator, and or amplifier to stabilize.
|
Measure
|
The minimum ammount of measurements in the measure loop. Continues to loop even if power level is stable.
|
Wait Time
|
The wait time before each measurement is taken, during the measurement-loop.
|
Max. Difference
|
If the measured power between each measurement in the measurement-loop, is less than this value. The measured power is stable. The loop will stop.
|
Max. Measure
|
The maximum ammount of measure times, for the measurement-loop
|
Both for the leveling, and the monitoring these settings can be set individually.
The leveling is used, whenever RadiMation® is measuring the power, to regulate the signal generator.
If the powermeter is only used as input, without regulating on it, then the monitoring settings are used.
More detailed information on how to achieve a fast and stable power measurement is available in Application Note 106: Powermeter parameter configuration.