RadiMation Error: 805306941

From RadiWiki
Revision as of 17:44, 16 February 2021 by Joro (talk | contribs) (Created page with "= GPIB system driver error (EDVR): Ethernet communication problem = A RadiMation device driver is trying to control a GPIB device through a GPIB board. However an error is...")
(diff) ← Older revision | Latest revision (diff) | Newer revision → (diff)
Jump to navigation Jump to search

GPIB system driver error (EDVR): Ethernet communication problem[edit]

A RadiMation device driver is trying to control a GPIB device through a GPIB board. However an error is raised by the GPIB system driver that controls that GPIB board. The driver has detected that there is an ethernet communication problem with the GPIB Ethernet device.

Solution[edit]

When a National Instruments GPIB board is used, the configuration of the GPIB board can be checked and changed in the NI-MAX software.

  • Check if the GPIB Ethernet device is still present and available in the NI-MAX software.