RadiMation Error: 536871970

From RadiWiki
Revision as of 15:22, 15 January 2015 by Joro (talk | contribs) (Created page with '= Too many datapoints = An emission test is configured, where the step frequency is too small. The number of points can be calculated by dividing the frequency span by the step f…')
(diff) ← Older revision | Latest revision (diff) | Newer revision → (diff)
Jump to navigation Jump to search

Too many datapoints[edit]

An emission test is configured, where the step frequency is too small. The number of points can be calculated by dividing the frequency span by the step frequency. The result is the number of steps that should be measured by the scanning receiver in a single sweep. The maximum number of data points in a single sweep is limited to 1.000.000 in RadiMation®. If the calculated number of required data points is higher, the 'Too many datapoints' error will be generated.

Solution[edit]

  • Increase the step frequency of the emission test. A step frequency that is half of the selected RBW filter is very often a good value.
  • Decrease the span of the emission measurement. Another possibility is to split the to be measured frequency range in different bands.