Release Notes/Solved items Version 2020.1
Jump to navigation
Jump to search
RadiMation version 2020.1.1[edit]
- #10181: RSRESW8 - Receiver (ID1703) Input 1 [LAN]: More points received from the receiver/analyzer then calculated.
- #10647: ErrorReport: "DARE!! Instruments RGN2400A: The requested carrier level is higher than the maximum ( 13 dBm). Limited the set carrier level to 13 dBm. "
- #10648: Include the RadiMation version number in the text of the e-mail of the error-report
- #10655: Loading an older TSF file, shows the test configuration in readonly mode
- #10656: Include a menu entry in the help menu with a link to the available application notes on the RadiWiki website
- #10711: update darelang to accept devexpress controls
- #10735: Not possible to select 3 dB filters on ESPI3.
- #10750: Ampersands are shown in the new ribbon menu
- #10753: Drivers: MEB CDN's
- #10767: R&S FPL 1007 - Screen is not switched on by default. Have to turn it on manually.
- #10769: R&S FPL 1007 - The set value of attenuator is not set in the spectrum analyser. During measurements value is set to 0 dB.
- #10770: R&S FPL 1007: non CISPR detectors chosen for 200 hz and 1Mhz.