Release Notes/Solved items Version 5.3
Jump to navigation
Jump to search
RadiMation version 5.3.21[edit]
- #2113: Keyword CALIBRATION FILE does not work with RI subst test reporting
RadiMation version 5.3.20[edit]
- #2971: Setting the puls setting to 0.5 uF is not remembered in the test.
- #3044: RI multiband ALL graphs key word does not work.
- #3133: dBpT/uV factor is not taken into account in the magnetic field test
RadiMation version 5.3.19[edit]
- #2930: limit line does not show during the merge function
- #2963: Crash when the R&S ESHS30 receiver is initialized
- #3012: the mode of operation is not inserted in the multi band emission test print out
- #3065: Why isn't average field data supplied during field probe calibration (X,Y & Z axis measurements are) at D.A.R.E!! Calibrations? What does Radimation require for calibration data?
- #3066: driver for LeCroy 9310AL
- #3069: Control of the turntable is not working correctly
- #3071: On emissions graph it reads dBuV and not dBuV/m depending on radiated or conducted
RadiMation version 5.3.18[edit]
- #2790: Wrongly measured reflected power during BCI-(fixed current)-tests
- #2873: Saving a TSF with a ':' in the name is raising errors
- #2878: Driver: Amplifiers
- #2886: Current is not displayed correct
- #2889: R&S ESPI: Data out of range during BCI Closed loop
- #2921: Combined Graph Report generator code is not replaced if the first mentioned graph is not available
- #2960: GPIB screen of the inco 2000 does not show
- #3006: Cal file is giving warning in sequence
- #3022: Error when trying to configure the R&S NAP driver.
- #3029: The limit line is missing in the table
- #3033: Report generation is giving the graphs I need
- #3074: Appears to be CR line feed errors in the notes / comments column of the performed tests table in the EUT Window
- #3101: Several problems with the HP 4395A driver
- #3104: Continuation of a Multiband immunity tests, starts with a too high signal generator level, causing overtesting
- #3113: Temporary bug to test defect creation
- #3127: Error message during initialisation of ESIB26
- #3165: Internal modulation of the Agilent 83640B is not correctly controlled
- #3228: FP4080 driver is crashing with 5.3.18.
RadiMation version 5.3.17[edit]
- #1890: Limitation on x-times the calibrated power is not easy to configure in multiband immunity
- #2144: Show related limit lines during measurement of trace in multiband emission
- #2980: NB/BB discrimination during CE and RE cannot be turned off
- #3110: Graph of an AD Channel from a Multiband immunity test cannot be included in the report
RadiMation version 5.3.16[edit]
- #1292: Time stamping of immunity frequencies
- #1608: Multiband emission should check the valid frequency range of all equipment during initialiation of the equipment
- #1873: If the name of the graph line is starting with '-' then the graph menu shows a caption
- #2178: Partially performed emission scan is not saved
- #2229: Detect peaks on the average detector sweep in multiband emission
- #2669: Show a table in multiband immunity with the data that is shown in the graph
- #2828: Measurement with the average trace detector, looses information of an already performed measurement of the peak trace detector
- #2834: RadiTurn raises an error 'already in progress'
- #2841: Emission measurement with a different detector erases already measured data of another detector
- #2854: Improvements in the read-out of the software protection key
- #2870: In the RE Chamber TSF, the 'General info' and 'Reporting' buttons are always disabled
RadiMation version 5.3.15[edit]
- #2133: Intermediate retrieved measurement 'blocks' from a scanning receiver are not updated in multiband graph
- #2332: When is the delete button active?
- #2339: Negative turntable angles are not allowed
- #2646: VZLAMP.dll is not installed during the set up
- #2677: adding device driver configuration in a protected config directory does not generate an error.
- #2784: Driver: CDN's for install Savonia
- #2831: Printing the graph from the multiband conducted immunity test raises an error
- #2836: Incorrect jig transfer factor is used for the calibration jig
- #2845: RadiSense driver does not work when used inside the radiCentre
- #2867: Power measurement crashes when the max power setting is above 10 times
- #2875: Agilent N5181A does not turn the Mod off
RadiMation version 5.3.14[edit]
- #2821: Driver crash when using Boonton 9200
- #2814: Driver: Amplifiers for 08D00011 Nato
- #2809: Report generator stops sooner then expected in excel report
- #2807: EUT specific information cannot be filled in for a newly created EUT file
- #2802: Test scripts do not report the amount of errors and warnings in the caption of the e-mail
- #2765: RADNI.DLL not supported anymore
- #2749: John deer text generator is crashing the software
- #2709: Frequency Offset of ESPC driver
- #2660: Legends of a graph line should only be shown if the graphline is really visible in the graph
- #2103: Upgrade Code Block Ide so we can use the --rebuild in the command prompt
- #2100: Upgrade wxWidgets library to use version 2.8.0
- #1829: During the measurement, only the Both-Peak-Max-Graph should be visible.
- #1687: Report generation slow (on hyperthreading/multiple CPU)
RadiMation version 5.3.13[edit]
- #2813: Coupler driver needed for Rockwell Collins
- #2801: Radiated immunity (single band) files are reported to be corrupted if they are opened directly after the test has been finished
- #2800: Graph lines of emission spectrum is not shown in generated report for multiband emission tests
- #2792: Software limits of antenna tower device drivers cannot be configured in the device driver
- #2785: R&S smiq driver is not waiting for the power to be leveled.
- #2782: Driver: ESD guns for install Savonia
- #2781: Driver Antenna for install
- #2771: multiple Solartron 7150 ad convertors in one
- #2757: Double click on the name of the engineer doesn't allow to edit the details
- #2748: removing amplifier correction file from signal input is not easy when the file is no longer available
- #2708: Driver: Agilent N9310A
- #2707: RadiPower Numbering is wrong
- #2676: Multiband with virtual devices, zero delay time internally, is taking more then 20 min to run
- #2353: Multiband emission is too slow
- #2227: Graph update is delaying the measurement.
- #1848: CPU usage is at the beginning of the test at 20-30% and after some steps at 100%
- #1836: A request to the user, if the software should really be closed before the program is closed.
- #1754: Multiband: Spectrum analyser specific settings can be changed when a (scanning) receiver is selected
- #1674: Emission test do not report out of range device like immunity test do
RadiMation version 5.3.12[edit]
- #2772: frequency checking in GTem test.
- #2743: Report generator doesn't replace code in first row of table
- #2724: Report generator should replace the complete paragraph if ||InsertFile(...)|| is used
- #2718: Report generator code is not replaced, if the contents of the document are changed
- #2651: reading back the information of the harmonic test is crashing due to file synchronization
RadiMation version 5.3.11[edit]
- #2760: Conversion to and from pF is not correct
- #2677: adding device driver configuration in a protected config directory does not generate an error.
RadiMation version 5.3.8[edit]
- #2264: Driver: Werlatone C1795 not supported
RadiMation version 5.3.7[edit]
- #2736: Amplifier Signal Generator protection is not triggered
- #2729: FM 5004 does not remember the GPIB address
- #2722: Correction file is added to the field instead of multiplied in multiband RI test
- #2698: Correction file is not correctly using the selected precision of the unit
- #2485: opening and closing the config window of the new GNU driver for the ESI is draining resource.
- #2484: memory leak with lisn test and new driver ESI
RadiMation version 5.3.5[edit]
- #2705: Just created calibration files cannot be opened anymore by the same software version (5.3.3)
- #1977: Progress calculation of Immunity test is going to 101%
- #1643: Testlevel unit is not shown in RI Mil-Std Calibration when a new TSF is created
RadiMation version 5.3.4[edit]
- #2693: Title head is not the same as the filename
- #2691: Sql Filter on Test site some times works and sometimes it does not
- #2686: A python script to fastly change unit settings in all TSF files
- #2685: Filtering of columns in a report generator table using 'SELECT ...... FROM <CODE>' is not working if requested columns are not in the generated table
- #2684: antenne meting does not support RECEIVED POWER_RAW in the report.
- #2678: Multiband Immunity needs to have an antenna tower and turntable
- #2671: PM window multi band does not update the time correctly
- #2662: Report generator reports that peak conservation is used during pulse modulation
- #2661: Report generator doesn't correctly output used frequency step if a frequency list is used
- #2653: shutting down the analyser before stopping the test holds makes saving of the data impossible
- #1605: Multiband emission needs to have an antenna tower device in the testsite
RadiMation version 5.3.2[edit]
- #2640: CE Multiband test is giving ‘horizontal’ instead of the lisn line
- #2633: Button in the environmental data window to use the same value as the last test for all the parameters
- #2614: Question: Which value for ALC is set inside the Kalmus 757LC?
- #2613: Report generation is slow when codes are included that are not present in the data file
- #2612: ||EUT|ENVIRONMENT|| is not replaced with the data in a report
- #2605: driver: Frankonia stripline, 90Ohm, 1GHz SR90/1000
- #2585: List of selected equipment is not sorted by default
- #2578: Radimation stops when not used for more then 10 minutes
- #2576: URV5 is returning measurement values in W instead of dBm
- #2572: Multiband immunity should ask on which location it should continue if automatic mode is enabled
- #2569: G-tem measurement is giving overflow error
- #2555: Driver ESN is not set up correctly
- #2549: Remove support for exotic report generator editors
- #2543: The port type selected in the cable configuration is not shown in the EUT window
- #2540: Crash when environmental data button is clicked in the TSF that is included in a sequence
- #2537: What is the usage of "Check Extension" in the report generator
- #2534: Typing error in manual mode window
- #2531: AR PM 2002 gives an 'Unknown Error' after zeroing
- #2530: driver: Schaefer Controller (Turn Table + changing polarisation)
- #2528: Environmental Data is some times not asked before testing
- #2523: Forward powermeter is not included in test initialization of a multiband immunity test
- #2518: Multiband frequency list is not taking the unit of the frequency column into account
- #2497: trying esi 40 driver on esi 26 is crashing the software.
- #2495: R&S ESI series analyzers hold the GPIB bus, so a GPIB delay should be used between GPIB calls of different device drivers
- #2490: Environmental data is empty when test results are reopened
- #2481: driver: NI-6221 DAQ card
- #2480: driver: NI-PCI-6014 Analog DAQ
- #2444: Include A/m as unit for magnetic flux density
- #2441: Measure more then 100 sweeps
- #2439: Possibility to select multiple cables/attenuations for a cable in the test-site
- #2322: Multiple correction files for RE chamber test
- #1837: Measure more then 100 peaks.
- #1444: Perform Immunity test from high to low frequency (Reverse frequency sweeping)
- #663: Turntable support in Radiated Immunity
- #484: Frequency steps per decade or frequency steps per octave not possible
- #285: Advanced Data Shower om alle data in de Test files te tonen
RadiMation version 5.3.1[edit]
- #2637: Driver PMM 8053A
- #2519: CE current probe test is hanging after the taking a sweep with a receiver
- #2517: Correction file limits values in frequency column to 40000 even if it is 40000 kHz
RadiMation version 5.3.0[edit]
- #2559: Reflected power limited when cable to antenna is inserted during BCI testing
- #2547: CAPL Interface
- #2487: Testing over network is slow when test become larger
- #2448: Driver of HI6105 is generating error 1 (ERROR_INVALID_FUNCTION)
- #2435: Start of test Initialization event is not sent anymore to switchmatrix device drivers
- #2430: Graphs cannot be customized anymore
- #2429: During the initialization of a window a strange (white) graph is seen
- #2428: Error that the turntable cannot be found in a testsite, if the starting angle is different from 0
- #2426: Software crash when loading TSF file in version 5.1.24
- #2424: Saving an EUT file to a read-only directory crashes RadiMation
- #2423: Power leveling in a multiband immunity test is not working
- #2418: Report generator code for advanced immunity data table
- #2417: Excel report generator should be able to use test specific templates
- #2410: EUT Angle offset is not transferred when starting a test
- #2396: Check the driver by pushing a button on the configuration screen.
- #2391: Storing the Drivers version in the test file
- #2390: Closing the eut window too fast kill the application
- #2380: Last test is not inserted according to sorting
- #2371: Radiated immunity fixed power information window does not contain a reporting button
- #2214: Single Radimation controlling the device driver system
- #2174: External test results
- #2141: Provide a link to the D.A.R.E!! Knowledge Base for detected errors
- #2104: Report window is too small to work with.
- #1965: RadiMation does not start if an old process is still running in the background
- #1736: Device driver advanced screen receives wrong hWnd
RadiMation version 5.3[edit]
- #2689: Tektronix TDS is missing detection if the marker is set correctly.
- #2670: Driver: AR150A250
- #2667: Driver: RadiMate with dBV unit
- #2659: Driver: R & S SMB 100A
- #2655: creating a new test configuration without a test site is crashing the software in amplifier compression test
- #2652: Extended frequency range for Rolf Heine LISN LN-KFZ20
- #2632: Driver: Agilent Model N5181A-506 (250 kHz – 6 GHz)
- #2624: Drivers: Injection devices and current probes and preamps
- #2623: Drivers: LISN drivers
- #2621: Drivers: Amplifier drivers
- #2620: Drivers: Antenna drivers
- #2618: FSEB 30 is not detected correctly
- #2617: driver: AR 5000A250
- #2608: driver: Frankonia FC 02 Positioning Controller
- #2592: HI4422 device driver is crashing when the advanced configuration window is closed
- #2546: Display protocol driver
- #2504: driver: R&S SMA100A signal generator
- #2493: Create external PM option for HP 8642B and HP 8673B
- #2477: anritsu analyser is giving invalid information when zooming in
- #2471: RadiCentre Multi Switcher - Not all bytes are written to the GPIB Board
- #2470: network analyser review window is not loaded in build version 5.3.909, keeping internal windows open
- #2464: Sequence file test does not allow removing of a test.
- #2463: Saving eut file can react as save as with same file name
- #2459: Description window does not match the name of the device
- #2458: inn co turn table stops turning during the test
- #2442: Device driver for agilent 83630A/83640A
- #2437: Driver: Kalmus Model 757C (via IF-488 interface)
- #2427: Driver: Bonn Elektronik BLWA 2010-50 & BSA 0122-100
- #2420: FSEB driver is giving error during start up option not available
- #2403: Empty Window message when starting RadiMation whed drives are gone.
- #2401: HP3478A used as a DC voltmeter gives 30% of actual value when using the recommended value from the manual
- #2394: Build scripts generate warnings when building the shared libs
- #2393: Manufacture information is cleared when pressing the X in the top right
- #2392: Add auto power on option in driver for FM7004
- #2389: fsh 6 driver generates overflow
- #2387: Driver: Tektronix TDS748C
- #2386: Driver: keithley 2700
- #2385: Driver: Lisn Schwarzbeck NSLK8126
- #2384: Pre amplifier driver Miteq AFSX 4...
- #2383: RadiSense driver does not support filter settings
- #2378: RadiMation Script fails because there are still 3 windows open message , but no window are displayed
- #2375: Font bottom of the graph is not the same as on the left
- #2324: GPIB driver of Schaffner CPW9670 does not work correctly while the rs232 does work correctly
- #2267: Driver: NRP-Z91 with USB Adapter NRP-Z4
- #2247: ESI driver is having problems when the observing time is equal to detector time
- #2188: Crash with esi on PML with RE multiband
- #2081: Messagebox without text pops up
- #2013: Customer database drivers cannot load a MS Access 2000 (and newer) database
- #2012: In MS Access customer database driver it is not clear if the database is correctly loaded
- #1779: Limit line configuration does not allow adding item at the top
- #1070: No external FM modulation possible
- #1066: No external pulse modulation capability