Difference between revisions of "Configurable AD Convertor"

From RadiWiki
Jump to: navigation, search
(1 Specify the communication type)
(How to configure a Configurable AD Convertor)
Line 11: Line 11:
 
<!-- Include a space to get a newline!! --> <!-- Beware of the space!! -->
 
<!-- Include a space to get a newline!! --> <!-- Beware of the space!! -->
  
[[Image:Configurable AD Converter Configuration Window.png]]
+
<div style="text-align: center; direction: ltr; margin-left: 0em;">[[Image:Configurable AD Converter Configuration Window.png]]</div>
  
 
== 1 Specify the communication type ==
 
== 1 Specify the communication type ==
Line 17: Line 17:
 
First the communication type must be configured.
 
First the communication type must be configured.
  
[[File:ConfigureCommunication.png]]
+
<div style="text-align: center; direction: ltr; margin-left: 0em;">[[File:ConfigureCommunication.png]]</div>
  
 
== 2 Configure the initialization of the device ==
 
== 2 Configure the initialization of the device ==

Revision as of 09:02, 6 October 2016

The Configurable AD Convertor device driver is a AD Converter which is supported by RadiMation®.  

How to configure a Configurable AD Convertor

Configurable AD Converter Configuration Window.png

1 Specify the communication type

First the communication type must be configured.

ConfigureCommunication.png

2 Configure the initialization of the device

ScreenElementDescription.png Reset The reset code that needs to be send to device. When left blank no command will be send.
ScreenElementDescription.png Init The Init code that needs to be send to device. When left blank no command will be send.
ScreenElementDescription.png Get Id The code that needs to be send to device to get the identification back. A common used SCPI command is *IDN?. When left blank no command will be send.



3 Configure the data readout of the device

ScreenElementDescription.png Trigger and Reading
ScreenElementDescription.png Trigger The command need to trigger the measurement. A common used SCPI is *TRG. When left blank no command will be send.
ScreenElementDescription.png Read Back In the read back the returned information is defined. This is different for every type of machine there for RadiMation® used regular expressions to understand the format. A good regular expression for finding the first number in scientific notation in a text is:
([+-]?[0-9]*\.?[0-9]+([eE][+-]?[0-9]+)?)

For more possibilities see the example section

ScreenElementDescription.png Trigger and Reading
ScreenElementDescription.png Minimum value The lowest value that can be measured with this device. The value must be the same as filled in the EUT window.
ScreenElementDescription.png Maximum value The highest value that can be measured with this device. The value must be the same as filled in the EUT window.


4 Configure multiple channels for 1 device

Examples

Received information
Regex
RadiMation Readout
Note
U_L_N 325
(-?[0-9.,Ee-]+)
325
Takes the first number.
I_L1;12.34
;(-?[0-9.,Ee-]+)
12.34
Takes the first number after ";".
THD_U_L1;;14,5
;;(-?[0-9.,Ee-]+)
14.5
Takes the first number after ";;".
AC_FREQ;Channel1;1.23E3
;*;(-?[0-9.,Ee-]+)
1230
Takes the first number after the appearance of a second ";".


Read backs can always be tested trough the "Validate Expression tool" ValidateExpressionTool.png