The left impedance is the signal generator which is generating enough power for 1 ampere.  
 
The left impedance is the signal generator which is generating enough power for 1 ampere.  
Line 56: Line 56:
  
 
{{note|This method is not a replacement for a real calibration as it may be performed by a none traceable device}}
 
{{note|This method is not a replacement for a real calibration as it may be performed by a none traceable device}}
 +
 +
[[Category:RadiMation]]

Latest revision as of 07:47, 31 July 2014

Theory

Current sensor Transfer impedance Prove.PNG

The left impedance is the signal generator which is generating enough power for 1 ampere.

This 1Amp. generates in the right impedance.

The current sensor has 1 ohm transfer impedance, this means 1 ampere generates 1 Volt on the measuring part below.

The power in the lower 50 ohm impedance is

So 1 ohm:

Reference measurement

Current sensor Transfer impedance Ref measurement.PNG

Probe measurement

Current sensor Transfer impedance final measurement.PNG

Calculation

and in dBm.


Example

On 10 MHz we have the following information:

  • Calibration: 0 dBm.
  • Measurement: -27,96 dBm.

So:

Verification

Current sensor Transfer impedance example.png

The left impedance is the signal generator which is generating enough power for 1 ampere.

This 1 Ampere generates in the right impedance.

The current sensor has 2 ohm transfer impedance, this means 1 ampere generates 2 Volt on the measuring part below.

The power in the lower 50 ohm impedance is

So 2 ohm:

The difference to a 1 Ohm impedance is

Conclusion

Correction to a 1 ohm impedance is


Information.png
Note: This method is not a replacement for a real calibration as it may be performed by a none traceable device