Difference between revisions of "RadiMation Application Note 154"
(→Configuration of the test sites) |
|||
Line 22: | Line 22: | ||
== Configuration of the test sites == | == Configuration of the test sites == | ||
+ | === Calibration === | ||
+ | [[File:CS114_Calibration.png]] | ||
+ | |||
+ | The configuration of the calibration test site should contain the following devices: | ||
+ | {|class="wikitable" | ||
+ | ! Device name || Tab in testsite configuration window || note | ||
+ | |- | ||
+ | | Signal generator || Devices 1 || The signal generator to use | ||
+ | |- | ||
+ | | Amplifier || Devices 1 || The amplifier to use | ||
+ | |- | ||
+ | | Coupler || Devices 1 || The coupler to use | ||
+ | |- | ||
+ | | Forward power meter || Devices 1 || The forward power meter to use | ||
+ | |- | ||
+ | | Reflected power meter || Devices 1 || The reflected power meter to use (optionally) | ||
+ | |- | ||
+ | | Sensor power meter || Devices 2 || The power meter or analyser to use for measuring the current | ||
+ | |- | ||
+ | | Current sensor || Devices 2 || The current sensor to use | ||
+ | |- | ||
+ | | Injection device || Devices 2 || The injection clamp to use | ||
+ | |- | ||
+ | | Jig || Devices 2 || The jig to use | ||
+ | |- | ||
+ | | Cable current -> power meter || Cables || Cable (1) with a correction file specified for the cable loss | ||
+ | |- | ||
+ | | Cable current -> power meter || Cables || Cable (2) with the specified loss of the used attenuator | ||
+ | |- | ||
+ | | Cable current -> power meter || Cables || Cable (3) with a correction file specified for the cable loss | ||
+ | |} | ||
+ | |||
+ | |||
+ | === Verification === | ||
+ | [[File:CS114_Verification.png]] | ||
+ | === EUT Testing === | ||
+ | [[File:CS114_EUT_Testing.png]] | ||
== Calibration procedure == | == Calibration procedure == |
Revision as of 14:46, 16 October 2023
How to perform a MIL-STD CS114, Conducted susceptibility, bulk cable injection test
This application note explains how the MIL-STD CS114, Conducted susceptibility, bulk cable injection test can be performed with RadiMation®
The exact requirements and test methods for the CS114 are specified in the MIL-STD-461.
Necessary equipment
The following devices are necessary to execute this test:
- Signal generator
- Amplifier
- Coupler
- Forward power meter
- Reflected power meter (optionally)
- Current sensor
- Sensor power meter / Analyser
- 50 Ohm load
- 50 Ohm attenuator
- LISN
- Current injection device
- Current injection calibration jig (Calibration fixture)
- Cable drivers with corrections
Configuration of the test sites
Calibration
The configuration of the calibration test site should contain the following devices:
Device name | Tab in testsite configuration window | note |
---|---|---|
Signal generator | Devices 1 | The signal generator to use |
Amplifier | Devices 1 | The amplifier to use |
Coupler | Devices 1 | The coupler to use |
Forward power meter | Devices 1 | The forward power meter to use |
Reflected power meter | Devices 1 | The reflected power meter to use (optionally) |
Sensor power meter | Devices 2 | The power meter or analyser to use for measuring the current |
Current sensor | Devices 2 | The current sensor to use |
Injection device | Devices 2 | The injection clamp to use |
Jig | Devices 2 | The jig to use |
Cable current -> power meter | Cables | Cable (1) with a correction file specified for the cable loss |
Cable current -> power meter | Cables | Cable (2) with the specified loss of the used attenuator |
Cable current -> power meter | Cables | Cable (3) with a correction file specified for the cable loss |