= The 'Variable' test-level is not supported. Use a 'Constant' test-level. =
 
= The 'Variable' test-level is not supported. Use a 'Constant' test-level. =
When an antenna scan calibration is performed, this can be performed with the 'signal generator method' or the 'network analyzer method'.  
+
When an antenna scan calibration is performed, this can be performed with the 'signal generator method' or the 'network analyser method'.  
* Signal generator method: will control the signal generator to regulate the forward power to the desired power level, and measure with the spectrum analyzer on the receiving antenna.
+
* Signal generator method: will control the signal generator to regulate the forward power to the desired power level, and measure with the spectrum analyser on the receiving antenna.
* Network analyzer method: will only use the network analyzer to generate a power level on one port, and measure the received signal on the receiving antenna with the other port.
+
* Network analyser method: will only use the network analyser to generate a power level on one port, and measure the received signal on the receiving antenna with the other port.
If a network analyzer is present in the selected testsite the 'Network analyzer method' will be used, otherwise the 'Signal generator method' will be used.
+
If a network analyser is present in the selected testsite the 'Network analyser method' will be used, otherwise the 'Signal generator method' will be used.
  
This error indicates that a variable test-level is selected in the TSF, in combination with the 'Network analyzer method'. When the 'Network analyzer method' is used, it is only allowed to use the 'Constant' 'test-level'.
+
This error indicates that a variable test-level is selected in the TSF, in combination with the 'Network analyser method'. When the 'Network analyser method' is used, it is only allowed to use the 'Constant' 'test-level'.
  
 
== Solution ==
 
== Solution ==
 
* Select the 'Constant' 'Test-level'.
 
* Select the 'Constant' 'Test-level'.
* Use the 'Signal generator method', by removing the 'Network analyzer' device driver from the selected testsite.
+
* Use the 'Signal generator method', by removing the 'Network analyser' device driver from the selected testsite.
  
 
[[Category:RadiMation]]
 
[[Category:RadiMation]]
 
[[Category:RadiMation Error]]
 
[[Category:RadiMation Error]]

Latest revision as of 09:19, 3 January 2020

The 'Variable' test-level is not supported. Use a 'Constant' test-level.

When an antenna scan calibration is performed, this can be performed with the 'signal generator method' or the 'network analyser method'.

  • Signal generator method: will control the signal generator to regulate the forward power to the desired power level, and measure with the spectrum analyser on the receiving antenna.
  • Network analyser method: will only use the network analyser to generate a power level on one port, and measure the received signal on the receiving antenna with the other port.

If a network analyser is present in the selected testsite the 'Network analyser method' will be used, otherwise the 'Signal generator method' will be used.

This error indicates that a variable test-level is selected in the TSF, in combination with the 'Network analyser method'. When the 'Network analyser method' is used, it is only allowed to use the 'Constant' 'test-level'.

Solution

  • Select the 'Constant' 'Test-level'.
  • Use the 'Signal generator method', by removing the 'Network analyser' device driver from the selected testsite.