RadiMation Error: 805334180

From RadiWiki
Revision as of 08:14, 8 September 2016 by Joro (talk | contribs) (Created page with '= The requested detector type is not supported = During the control of the Aaronia Spectran analyzer, it was not possible to set the desired detector. The Aaronia Spectran on…')
(diff) ← Older revision | Latest revision (diff) | Newer revision → (diff)
Jump to: navigation, search

The requested detector type is not supported

During the control of the Aaronia Spectran analyzer, it was not possible to set the desired detector. The Aaronia Spectran only supports the 'PK' and 'RMS' detectors. The 'Average' and 'Quasi-Peak' detectors are not supported.

Solution

Possible reasons for this error can be:

  • Do not select the 'Average' or the 'Quasi-Peak' detector.
  • A wrong configuration of one of the emission measurement parameters. Try to change the parameters to more often used values.

If this problem remains after manually checking the above options, further investigation by DARE!! Instruments is necessary to find the real cause of this problem.