Difference between revisions of "Rohde & Schwarz NRP"

From RadiWiki
Jump to: navigation, search
 
Line 16: Line 16:
  
 
It is recommended that the frequency send to the head is done in Hz instead of MHz. The head may not respond to MHz for the frequency correction.
 
It is recommended that the frequency send to the head is done in Hz instead of MHz. The head may not respond to MHz for the frequency correction.
 +
 +
== Installation and remote communication ==
 +
For more information about these subjects see: [[Rohde & Schwarz NRP-Z91|Rohde & Schwarz NRP-Z91]]
  
 
== Links ==
 
== Links ==
 
* [http://www2.rohde-schwarz.com/product/nrp.html  Website]
 
* [http://www2.rohde-schwarz.com/product/nrp.html  Website]

Latest revision as of 09:33, 1 July 2015

The Rohde & Schwarz NRP device driver is a Power Meter which is supported by RadiMation®.

Manual

  • The NRP manual can be found here

Frequency range

In the manual page 6.9 of the remote control, there is a description regarding the frequency area of correction of the frequency response. This goes from 10.0e6 to 40.0e9 Hz for the NRP-Z55. However the frequency range may vary, depending on the head.

For the NRP-Z55, it does measure below 10 MHz, but because of wide band width of these measuring heads, no level correction for these frequencies are done.

It is recommended that the frequency send to the head is done in Hz instead of MHz. The head may not respond to MHz for the frequency correction.

Installation and remote communication

For more information about these subjects see: Rohde & Schwarz NRP-Z91

Links