Release Notes 2019.1
Introduction[edit]
We are proud to deliver you Version 2019.1 of our RadiMation® EMC Test software. This 2019.1 version is the result of several major improvements that we have done.
This letter contains all the information that is needed before your system is upgraded. If you have any additional questions or remarks please let us know through your reseller, or contact radimation-support@raditeq.com.
Installation[edit]
Backup[edit]
When RadiMation® files are opened, changed and saved in a newer version of RadiMation®, it is possible that it will saved in a newer file format, which cannot be opened anymore in an older version of RadiMation®. To prevent that changed TSF, LLF, COR, SEQ or EUT files cannot be opened anymore in an older RadiMation® version, we strongly recommend to make a backup. Before you install the latest version of RadiMation®, please make sure to saveguard all your critical data.
Make a back up of (at least) the following files:
- Correction files
- TSF files
- Sequence files
- Limit Line files
- Device driver configuration files
Multi version installation[edit]
It is possible to install different RadiMation® versions on the same PC.
This functionality can be used to verify the correct functioning of the RadiMation® software before it is used for EMC testing. Simply put, the functioning of the new version can be checked by comparing it with the older version. This verification can for example be required by internal quality assurance standards. During this verification process, the installation of the older RadiMation® version can still be used for all EMC tests.
Be advised that:
- It is not possible to run two (or more) different versions at the same time without risking a software crash. Close other instances of RadiMation® before starting another version!
- Raditeq cannot be held accountable if information is lost because two different versions of RadiMation® are used at the same time.
- The user needs to have the rights as described in User rights and installation.
Installation[edit]
Installing software updates[edit]
Software updates can be distributed on CD or downloaded from the RadiMation® Download page.
To install the RadiMation® update from a CD, please execute the ‘SETUP.EXE’ startup program. The setup program will install the new RadiMation® version at the same location as the previous version, but in a separate installation folder. This will allow you to use the previously installed RadiMation® version, as well as the new one.
The setup program will detect that a previous version of RadiMation® has already been installed on your PC. A back up of this older version will automatically be copied to the backup directory. All the settings of the older version will also be used for the newly installed RadiMation®. If any problems should occur with the new software, the older version can be recovered.
Installing from a network directory[edit]
If the already installed RadiMation® version was installed from a network directory, the new setup file can be copied to the <Setup> sub-directory of the <NetworkDir>.
The new RadiMation® version can then be installed on each computer from this directory.
Updating device drivers[edit]
After the new RadiMation® version is installed, we strongly recommend that you install the latest device drivers as well. These device drivers can be installed by starting the ‘DRIVERS.EXE’ program that is located on the delivered CD.
The ‘DRIVERS.EXE’ program will detect the different RadiMation® versions installed on your PC. A dialog will start, asking you to select the RadiMation® version that needs updated device drivers. Select the latest version of RadiMation® that was just installed.
You have successfully obtained the latest released version of RadiMation® when both the software and device drivers have been updated.
Verification and Validation[edit]
Every new RadiMation® version is tested extensively before it is released. Raditeq validates these new versions on three ‘levels’. The fourth level of the validation is to be done by the end-user.
This quality management system follows the ISO/IEC 17025 standard (“General requirements for the competence of testing and calibration laboratories”).
1) Automatic knob pushing program[edit]
The ‘automatic knob pushing program’ simulates an end-user and is run on each RadiMation® version. The program performs all the actions that a user could do, such as entering information, executing tests, reviewing test results and generating reports.
Performing the first validation level is a responsibility of Raditeq.
2) Validation and usage in a test lab[edit]
Once the new RadiMation® version has passed the first validation level, it progresses to the second validation level performed by Raditeq.
During this validation the RadiMation® software is installed in a real test lab, to perform measurements on a 'known' EUT. These results are then compared to manual measurements to verify the quality of the software. The results are also compared with the measurement results of previous software versions.
Performing the second validation level is a responsibility of Raditeq.
3) Validation and usage in a test house[edit]
When the new RadiMation® version has passed the second validation level, it progresses to the third validation level performed by Kiwa Dare Services.
This validation level consists of two stages. In the first stage the software is used to perform measurements on a 'known' EUT. These results are then compared to manual measurements to verify the quality of the software. When the new software version has passed the first stage, it progresses to the second stage where it is installed and used by Kiwa Dare Services for at least two weeks. Any issues that are reported during this period, will be investigated and resolved. An updated version of RadiMation® is created if needed.
Performing the third validation level is a responsibility of Kiwa Dare Services.
When this validation is succesful, the RadiMation® version will be officially released, and will become available for all end-users.
4) Validation by end-users[edit]
Once the new RadiMation® version has passed the third validation level, it progresses to the fourth (and last) validation level, which should be performed by the end-users of RadiMation®.
When end-users receives a new RadiMation® version they should perform a validation of the software in their own test setup. The parts that should be validated by the end-user are:
- Correct control of the used measurement equipment
- Validation of the software according to internal quality standards
- Possible verification and comparison of measurements on a 'known' EUT
We strongly advise to only start to use RadiMation® after this validation is performed.
Performing the fourth validation level is a responsibility of the end-user.
Solved items[edit]
See below the list of all the issues that have been solved in RadiMation® version 2019.1 since version 2018.1
RadiMation version 2019.1.10 (BETA release)[edit]
- #10112: Radimation crash when opening the advanced dialog of a horn-antenna driver for a second time.
- #10357: Maturo NCD tower and turntable
- #10366: Driver: Receiver/Analyser Tektronix RSA6114A
- #10390: Driver: Tektronix TDS5054B AD Converter Driver
- #10399: EM-TEST UCS500N7 - Surge test is not working
- #10399: EM-TEST UCS500N7 - Surge test is not working
- #10417: Eventlog is showing an additional column 'Setting ID' with: 'DARE.RadiMat.clsDataIdentifierSetting'
- #10441: Polarity of test generator
- #10422: DARE!! Instruments RPR2006C: Timeout expired before operation completed.
- #10494: Driver: Pearson Model 101
- #10524: Driver: DARE!! Instruments RGN2006A
- #10531: MK Messtechnik - Initialization before AND After each test
- #10537: SMB 100A: Specific System Error -222, Data Out of Range
- #10541: Driver: DARE!! Instruments RGN2400A
- #10570: RI Substitution Single Band - modulation: Pulse states"With Voltage conservation"
- #10573: Driver: RadiGen 2400A changed commands.
- #10576: Driver for monitor input for the NSG4070
- #10578: EMTEST UCS500N7 - Surge test is not working when controlled remote
- #10582: Rohde & Schwarz ESI 40 Input 2: Specific System Error: -222,"Data out of range; SCAN1:BAND:RES 1 KHZ
- #10585: DARE!! Instruments RSW1021N: Wrong command Accepted use of data: True User
- #10596: The enhanced status window is popping up during a band switch.
- #10598: ||SELECT [Frequency (MHz)], [RBW: 9 kHz, Neutral Max Peak (dBµV)] FROM RBW: 9 kHz, Neutral Max Peak TABLE|| does not resolve.
- #10606: Driver: Wavecontrol SMP2
- #10608: Driver: R&S BBL200
- #10611: CI- When starting test we get the error "Teseq NSG 4070 Driver function not implemented"
- #10612: Driver: R&S USP220 switch matrix with module B107
- #10616: Rohde & Schwarz SMB 100A: Specific System Error: -222,"Data out of range;
- #10583: SMB 100A: Specific System Error -222, Data Out of Range: SOURCE: POWER 40.3894 DBM
- #10626: Error at SMF100A command in our driver
- #10627: PMM EP601 Not using selected filter settings
- #10630: Keysight technologies N5172B: Specific system error: -410, "Query INTERRUPTED"
RadiMation version 2019.1.9[edit]
- #8314: Communication is getting out of sync using a Configurable Antenna Tower driver
- #9740: Clarification BCI test limits to early on current limit
- #9847: RBW changed, radimation halted.
- #9935: Interlock errors are raised by an AR 5000A250A
- #10099: Device specific setting in the RPR3006W for 'Edge Fall Time'
- #10113: Mini-Circuits PWR-SEN-6LRMS-RC
- #10171: Polariser & antenna tower driver RAT1001B for use with RadiPol RPL2010A
- #10172: EMCO2090: Turntable - No notification when using a turntable position higher when the limit for he EMCO2090 is set at 180 degrees.
- #10242: Driver: TESEQ RF Switch RFB 6000.
- #10257: Object reference not set to an instance of an object.
- #10275: Error with FM modulation.
- #10334: Hewlett Packard 34401A (Multiple)- Device Not Connected.
- #10344: When I zoom in the limit line is moving.
- #10349: How to frequency range PMM 6630?
- #10358: Keysight Technologies N9322C: Specific System Error: -222,"Data out of range;BAND:TYPE CSP2
- #10362: When we want to print Start-Stop frequency we only get the result from the last band.
- #10375: Radimation crashes when checking the driver for R&S NRX.
- #10384: Add the Tracking generator option to ESR 7 & ESR 3.
- #10392: Driver: Keysight 34465A AD-Convertor.
- #10402: Spectrum Analyser 8563A - ERR 112 ???CMD??? Unrecognized Command.
- #10402: Spectrum Analyser 8563A - ERR 112 ???CMD??? Unrecognized Command.
- #10459: Der_Objektverweis_wurde_nicht_auf_eine_Objektinstance_festgelegt
- #10464: Error connecting FCU: Not able to connect over IP.
- #10500: DARE!! Instruments RGN6000B HF (USB): Setting the output raised an error.
- #10511: RadiSense 10 Upload User Calibration table
- #10513: Extend maturo fcu clamp mover driver with speed configuration.
- #10521: Checking the driver for R&S NRX indicates that the device is not connected
- #10591: NARDA 3006 checking device crashed Radimation when there was no battery present and the device runs on a adapter
- #10650: Connection to CARAD not working
RadiMation version 2019.1.8[edit]
- #9868: Driver: EUT Monitoring: HP 8903B : Audio Analyser
- #10049: Driver: EMC Partner Transient 3000
- #10186: CE: R&S ESPI3A RBW has changed from 9 Khz to 120 Khz
- #10258: When running a longevity test there was a communication error "No response is received from device". Not able to close radimation by pressing cancel needed to force quit to leave Radimation.
- #10307: RF Amplifier (ID1651) - Amp hangs in init.
- #10341: RE multiband receiver mode - 100 Khz set in software, 9 Khz set on analyser
- #10342: multiband receiver mode - 100 Khz set in software, 3 MHz set on analyser
- #10343: RadiMation Printout Error: "Befehl misslungen, Error: -2146824090".
- #10348: export polar plot max peak for peak not possible anymore in 2019.1.5
- #10354: Equipment table list is missing
- #10367: Driver: Eut monitoring LeCroy WaveRunner 620Zi
- #10368: Driver: EUT monitoring using P2 (Yoda)
- #10369: Driver: EUT Monitoring LeCroy HDO6054 LAN
- #10370: Driver: EUT Monitoring Bus System
- #10371: Driver: Amplifier Bonn BLMA 2060-50
- #10372: Driver: Amplifier BLWA 0825-160/100/70D
- #10373: Driver: Passive drivers
- #10382: Conducted emissie multi band measuring in analyser mode 120 K and 1 Mhz is selected while all bands set to 9 Khz
- #10396: Possibility to change the septum height for the calculated field in the configurable switchmatrix device driver
- #10402: Spectrum Analyser 8563A - ERR 112 ???CMD??? Unrecognized Command.
- #10409: Selection of capacitive coupling clamp on compact nx5 does not work
- #10411: Regulation of power is not done within the desired tolerance
- #10431: Fixed issue with ESI based driver. Not possible to select more then one trace.
- #10464: Changed timing on CM NP GO for FCU which prevented controlling the clamp
- #10551: Added comments in code regarding the dependencey on RSSINS32.dll
- #10649: Emission - Measure time in Multiband and Singleband
RadiMation version 2019.1.7[edit]
- #8633: System.OutOfMemoryException was thrown.
- #9888: Agilent 8753: Specific System Error: 50 , "CAUTION: TOO MANY SEGMENTS OR POINTS"".
- #9998: N9322C analyser gives BAND:AUTO ON error message
- #10073: -150dBm read out from PWR-6Ghz power sensor
- #10134: Keysight Technologies N9322C: Specific System Error: -222,"Data out of range;DET:TRAC1:FUNC POS;<Err>"\n)
- #10153: RE Manual Mode - N9322C error measuring above 1 GHZ - Keysight Technologies N9322C: Specific System Error: -222,"Data out of range;BAND:AUTO ON;<Err>"\n
- #10164: Frequency column is not using settings of unit in RI frequency table.
- #10168: Cannot find antenna tower in the currently active testsite.
- #10222: N9322C timeout
- #10224: Keysight Technologies N9322C: Specific System Error: -222,"Data out of range;:FREQ:STOP 6000000000 HZ
- #10237: Keysight Technologies N9322C: Specific System Error: -222,"Data out of range;BAND 100000
- #10255: Keysight Technologies N9322C: Specific System Error: -222,"Data out of range;:FREQ:STOP 6000000000 HZ
- #10294: Driver: R&S ZVB8 Network analyser
- #10298: Final Quasi Peak measurement of R&S ESRP is suddenly done in analyser mode which takes a very long time
- #10300: R&S ZVB takes an additional unnecessary sweep during the port calibration
- #10301: Allow to select External Reference for R&S ZVB8
- #10319: Rohde & Schwarz ESU 26 Input 1 (LAN): RSIB: Timeout has occurred.
- #10325: Rohde & Schwarz ESPI 3 Scanning Receiver: Impossible to get description string for this Error! Probably an unknown RadiMation Error. Code: 805309369 (0x30000BB9)
- #10332: Ability to use the R&S ESW with more than 32000 points
RadiMation version 2019.1.6[edit]
- #7305: Communication with the TDEMI should be optimized for Continuous Measurement
- #9669: Extend Gauss Instruments TDEMI driver to allow selection of CISPR-Average or (old-)Average as Average detector
- #10043: EM-Test NX5: Selected line coupling is not possible.
- #10053: Exception has been thrown by the target of an invocation.
- #10054: AD converter limit lines not visible in time graph.
- #10098: Object reference not set to an instance of an object, when editing a non existing cable.
- #10105: powersensors not recognized HP438
- #10110: Make sure that using |||| in the report generator is correctly handled and does not trigger a infinite loop.
- #10121: Voltage dips and interupts: Sequence not starting at number 1.
- #10124: VDI - When a test is closed it is not possible to change the settings Test failed/passed
- #10125: Sequence order disfunction.
- #10126: R&S ESR7 - Preamplifier gets always inactive when starting a measurement
- #10127: RI: Band is shown in report as 43,567 Ghz while it was 1Ghz in test.
- #10141: EUT calibration-Attenuation/Gain: Limitlines are not exported.
- #10175: Internal frequency correction of DARE!! RadiSense 10 probe should by default be off
- #10176: Graphs: Setting a default for a line is not loaded correctly upon starting a new test.
- #10189: 17-03 Prana SX 120-100 low: Amplifier has not correctly set the desired range.
- #10199: Value cannot be null.
- #10214: Driver: EMSense 10
- #10232: Auto scale height graph.
- #10277: De objectverwijzing is niet op een exemplaar van een object ingesteld.
- #10278: Hewlett Packard 8563A: Not all bytes are written to the GPIB-Board within the timeout period (GPIB: EABO).
- #10279: Rohde & Schwarz ESR 26 Input 2 Scanning Receiver: Specific System Error: -222,"Data out of range; :SCAN1:STAR 0 HZ
- #10289: CI: Stepsize in the test does not mach the stepsize setting in 2019.1.5.
- #10398: Icon now correctly displays status pass/fail
RadiMation version 2019.1.5[edit]
- #8672: Limit line file printout shows no units in the table for Frequency and Value.
- #9259: HTTPS version of RADIWIKI site
- #9519: SMR 4503: Not all bytes are written to the GPIB-Board within the timeout period (GPIB: EABO)
- #9529: emtest UCS 00N: One or several values are limited by the simulator. The paremeters are set to a minimum
- #9888: Agilent 8753: Specific System Error: 50 , "CAUTION: TOO MANY SEGMENTS OR POINTS""
- #9925: Reporting in Excel not functioning well in 2019.1.2
- #9945: Driver: R&S NRX
- #9970: DARE!! Instruments RSS2010I: Device stream is not connected
- #9971: Schlöder SFT 1400: An unexpected unit code is returned. Is another Schloeder model connected?
- #10014: Radiated Emission Tower does not stop
- #10045: Activate HTTPS support for HelixALM website
- #10052: Keysight Technologies DSOX4154A
- #10082: Rounded frequency steps.
- #10099: Device specific setting in the RPR3006W for 'Edge Fall Time'
- #10109: "|| TEMPERATURE ||" exports wrong string
- #10115: Switch matrix sends scientific notation numbers
- #10119: EM-Test Compact NX5 VDIP - checkered flag not visible after test
- #10148: Pulsed immunity EFT Multiband - Export icons pass are not consistent.
- #10156: CI: Forward power reads 40 db too high
- #10159: RadiMation 2019.1.4 using a fixed min and max for Y axis does not work anymore.
- #10160: 1GHz radiated EMI: Amplitude scale not automatically adjusted.
- #10161: CI- 2019.1.4 does not output the time column in the report generator anymmore
- #10163: Gauss Instruments TDEMI: Invalid parameter value.
- #10169: When logging the results of the CW measurements in the multiband method, the line-wise assignment between frequency and signal power and / or output power is not correct! The same applies to the documentation of the Apparent Power
- #10170: The graphs for harmonic and compression power measurements are incorrectly formatted. This may be due to the incorrect designation of the frequency axis (Hz instead of MHz).
- #10174: Graphs: Fixed limits are scaling instead of being fixed.
- #10179: VDI Repetition time is cut when changing angle for UCS500
- #10187: Exporting Electrical field probe line as cal file not possible
- #10191: Rohde & Schwarz ESPI 3 Scanning Receiver: Specific System Error: -222,"Data out of range; :FREQ:STOP 9000 HZ"
- #10193: Index out of bounds when showing polar plot.
- #10194: Lock timeout on etherds
- #10195: Height Graph: Height plot: Max Peak for peak 1 not drawn in graph.
- #10233: Updated the manual to reflect the current UI for cables
- #10229: DARE!! Instruments RPR3006W: Wrong command
- #10230: Resolves IDN check issue with 30S1G3
- #10244: RadiMation not responding - freezes
- #10250: Saving TSF in directory crash, and TSF seemingly lost
- #10256: ESRP 7 Scan max min only usable in scanmode
- #10261: Analyser: Gauss TDEMI 9 Wrong RBW 9-150 kHz QP scan
- #10314: Some graphlines are displayed multiple times in the default Custom Graph window
- #10329: TDEMI M6+ is much slower with Radimation then doing the test manualy
- #10332: Ability to use the R&S ESW with more than 32000 points
- #10337: How can we directly control level of injection current in CI & How can we set the level of field strength in RI
RadiMation version 2019.1.4[edit]
- #7991: VDI Repetition time is cut when changing angle for PSF500
- #8086: R&S ESW26 abillity to set number of points on X-axis
- #9559: RF Amplifier (ID1344): Prana returned a warning state.
- #9774: CE No trace when using 200 Hz RBW and a 3 Khz VBW
- #10011: When opening an EUT file the directory should be set to the EUT directory.
- #10012: The File extension of the template and the output file do not match
- #10020: Collection was modified; enumeration operation may not execute.
- #10066: Compact NX5 surge testing halted
- #10069: Timeout time configurable ADC
- #10077: System calibration: Reference attenuation: Highest frequency is sometimes 150 instead of inline with the measurements
- #10081: Mini-Circuits SSG-6001RC generates a higher carrier level than the requested 0.0 dBm
- #10122: Marconi 2024 (ID1092) - Error 151: Query UNTERMINATED
- #10133: Keysight Technologies N9322C: EMC Option is not present in this device.
- #10140: Correct keywords for Pulsed Immunity SURGE MB?
- #10166: RE: TDEMI X6: Invalid parameter value.
RadiMation version 2019.1.3[edit]
- #8139: Updating consistency in code
- #8211: Object reference not set to an instance of an object: NoStacktrace.
- #8671: RI Fixed power Printout stepsize is incorrect when frequency list used.
- #8760: The CheckDevice check for Amplifier Research amplifiers using the P0/P1 command set is always successful
- #8828: Virtual/manual device driver for EFT
- #8868: VDI Test not running correctly
- #8898: RepGen .xlsx EUT attachment ||EUT|ATTACHMENT 1 CONTENTS||
- #8940: External modulation option for signal generator disappears with RadiMation Free
- #9072: Receiver (ID1691) Input 2: Specific System Error: -200,"Execution error; Function not available; ABORT
- #9083: RBW values are not working for ESU 8
- #9083: Unable to set device driver auto value for measure time, incombination with RBW value
- #9137: Driver: Mini-Circuits SSG-6001RC
- #9257: Update the contents of the RadiLog page on the RadiWiki
- #9391: Object reference not set to an instance of an object: Main Ticket.
- #9554: em test PFS500
- #9757: test interrupted and restarted with wrong RBW and wrong turntable sequence
- #9764: Crash when creating an EUT using the auto-EUT-naming feature.
- #9848: R&S ZNB device driver should not transmit measurement frequencies again when they are not changed
- #9859: EF01003 (FSU 26) LAN ??? 10.49.58.154: Specific System Error: -200,"Execution error;Function not available;BAND:AUTO ON""
- #9880: Radimation hangs during closing, when it is waiting on an errornous amplifier
- #9928: Amplifier Research 150W1000 (Remote controlled): Device Not Connected.
- #9937: Driver: RadiLink new relais
- #9941: Missing Radi Errors descriptions
- #9942: Incorrect frequency in error?
- #9945: Driver: R&S NRX
- #9948: Exception has been thrown by the target of an invocation, when EUT Monitoring channel is opened
- #9950: Controlling a device over HTTP causes unnecessary delays
- #9952: Rohde & Schwarz FSL 3: Specific System Error: -113,"Undefined header;INP:PRES:STAT ON"\n
- #9963: RadiMation crashed during setup AD converter
- #9979: CE: Virtual device - Device not ready
- #9986: Turntable go back to 0deg instead of continuing rotation to 360?? after interrupting test
- #9991: Change Canoe Switch matrix driver so that you can choose which application it starts
- #9992: No error messages are retrieved from Mini-circuits PWR-6GHS
- #9993: Antenna stays at the last position
- #10001: Constant timeout errors with AD-readout
- #10009: Driver: Schwarzbeck TK9420
- #10013: Driver: Anritsu MS2601A as powermeter
- #10018: Check for UNCAL after a Factory Reset of R&S Receiver
- #10023: AntennaTower: Does not return to base position after test in version 2018.2.9
- #10039: Unable to set RBW value
- #10040: Low band of RSS2010B is not measuring the fieldstrength correctly
- #10066: Compact NX5 surge testing halted
- #10068: Show all available ad channels instead of the first 8
RadiMation version 2019.1.2[edit]
- #7201: VDI: Reopening TSF still causes error with different location types.
- #8309: The magnetic field current is not shown in multiband immunity.
- #9686: Calibration method of the a multiband immunity test cannot be retrieved in the report.
- #9691: Reference level is not changed to 125 dBuV on a R&S ESU
- #9763: Index out of range when entering description in testlevel auto filter box.
- #9780: Antenna scan calibration in combination with network analyser should use S21
- #9841: Update Visual Studio C++ compiler to 15.9.5
- #9842: ErrorReport: "Spectrum Analyser (ID1691) Input 2: Specific System Error: -200,"Execution error; Function not available; FREQ:STOP:LINK START""
- #9860: Object reference not set to an instance of an object.
- #9901: Could not load file or assembly 'System.ValueTuple, ....' or one of its dependencies.
- #10051: em test VDS 200Q: Sting error: error in data-transmission
- #10093: CE-MB test: not the exact peak is chosen which results in a large measurement error
RadiMation version 2019.1.1[edit]
- #1924: Zooming in the graph is not possible in compression test
- #4370: TSF file is duplicated
- #4670: Building the 'TestSites' menu can take a long time
- #4777: Printing the angle plot of a Multiband emission test is printing the frequency graph
- #5754: Graph of UFA calculation starts at -1, but should start at 0
- #6529: Version numbers in the Device driver DLL's should use the build number that is used, also if adjusted to the SVN revision
- #8613: If auto-save is active, data of measurement should be saved to disk when automatic mode is left
- #7793: Compression Gain System Calibration representation of graph is vague
- #7794: Compression Gain System Calibration when saved it removes rest of suggested name description after any.
- #7795: Compression Gain Test default value of added compression point is invalid
- #7869: Y axis scaling in graph improvements
- #8021: When a Band has been selected the TSF is marked as being changed
- #8024: Magnifier in directory configuration is opening wrong folder or not opening the folder, but after multiple clicks it is opening the selection window
- #8148: Modification to Monitoring input channel sometimes reset the configured channel to 1
- #8149: Graph Customization Configuration of the global graph color settings outside a test is not updating the global settings
- #8158: Multiband Emission TSF configuration is not using the maximum number of selectable detectors
- #8220: Graph customization is not applied when a test is opened
- #8265: Confirmation message is default selecting wrong answer (No) instead of Yes
- #8268: Graph Customization is not updated
- #8272: General Information in a calibration file cannot be shown if the calibration file is readonly
- #8277: Disconnecting one VI of four connected RadiCentre devices is raising errors
- #8283: A 'Copy to Clipboard' action is marking a COR file as being changed
- #8290: System Calibration Attenuation / Gain Message shows message to connect the sensor power, but requires a forward powermeter in test equipment
- #8318: Impossible to remove graph line from different test data when the file has been closed
- #8319: Multiband shows an unexpected frequency graph line and table information which can't be removed
- #8374: Multiple PC's running the same confdvdr location gives error
- #8380: Selecting a field probe in a testsite takes more than 20 seconds
- #8546: Index was out of range. Must be non-negative and less than the size of the collection.
- #8554: Ability to show polar plot of field strength in Multiband Immunity test
- #8563: Not possible to select 1 point per row in UFA calculation
- #8564: ErrorReport: "Index was outside the bounds of the array.
- #8587: Changing graph customizations is not setting the changed state of the test
- #8678: Allow 1 port network analyzer calibration for the Agilent 8753
- #8733: Adding a multiline note field in the general information items of the eut is not possible
- #8757: Not possible to edit MB EFT test from the sequence window
- #8812: Wrong lines are drawn in the Multiband Immunity graph, when the same frequencies are tested often
- #8820: Ability to use Spectrum analyzer for harmonics measurements
- #8864: Graphs of gain compression calibration are not showing all the data
- #8866: No error is raised when the fieldprobe has a multiply correction file attached with values below 0
- #8881: Graphs of amplifier linearity calibration cannot be shown in Watts unit
- #8882: Better auto scaling for the y-axis of the amplifier linearity graphs
- #8887: RadiMation free Virtual powermeter
- #8892: Wrong title Harmonics calibration
- #8893: Harmonics test graphs does not use stored graph settings
- #8904: Suggested filename for the calibration is not correct if the TSF contains a .
- #8918: print graph shows temp document already being used by another process
- #8938: Perform antenna scan calibration using a network analyzer instead of signal generator and spectrum analyzer
- #8941: ctlEvents.cs is included twice and seems to be duplicated
- #8948: Not all calibration tests are listed in the add test function of the sequence.
- #8974: Selecting multiple bands with shift is not working as it should be
- #9013: Sort the names of the graph as they are shown below the 'Graphs' button
- #9025: Table tab does not display its measured values in a correct manor
- #9032: Harmonics show -1000dBm
- #9062: Measurement of Harmonics with 2 power meters
- #9082: RadiMation calibration expire date database error crash
- #9100: Object reference not set to an instance of an object: clsGraphElements.FilterShownGraphElements
- #9105: GetGlobalizedHashCode Object must be of type String
- #9125: Pressing update in graph config does not update the image.
- #9137: Driver: Mini-Circuits SSG-6001RC
- #9185: Unclear time logging behaviour Radimation
- #9186: Suggested filename of a calibration file is only 'm.cal' if the TSF description is: "80-1000 MHz, 10 V/m"
- #9188: Corrupt testfile, no data is visible in the graph
- #9218: Infinit loop graph elements being added for gain calibration test
- #9219: Cannot find the device driver - (Device is not on - Device driver is installed)
- #9232: Emission peak number not correct
- #9239: When starting an isotropy measurement the angle tab is not displayed
- #9242: Exception is shown when EUT Monitoring input is opened
- #9248: Readonly Network Analyzer system calibration still has some fields enabled
- #9250: Number of sweeps in TSF of Network Analyzer EUT Calibration is not correctly initialized
- #9267: Continue button does not work
- #9287: Test data disappeared
- #9288: PEAK TABLE keyword no longer works.
- #9291: Save error report standard suggests .log extension
- #9292: When there is no EUT there is no log file in the error report
- #9299: Polar graph displays a graph line that has it's visibility tuned off
- #9303: ErrorReport: "Radimation crashes suddenly after CE measurement is finished "
- #9304: Exception of type 'System.OutOfMemoryException' was thrown.
- #9310: Testdata again not visible.
- #9311: Test data lost / not visible
- #9315: Auto Backup of RadiLog
- #9353: Popup menu of available devices has no seperator between the actions and the devices
- #9356: Configurable AD Converter value always 0
- #9357: Driver: new RadiSense 10 models
- #9384: Index out of range: clsGraphElements.HideElement.
- #9427: Cancel button of Multiband Immunity TSF window is not working anymore
- #9429: Destination array was not long enough
- #9431: External FM modulation waveform is not saved
- #9445: Battery radisense error, amp did not stop generating 100 vm
- #9449: Object reference not set to an instance of an object: clsImpedanceTableCalculation.GetResistanceFromImpedanceTable(
- #9452: Burst, graph does not re-scale when the to measure time is < measured time
- #9453: Close RadiMation with FieldSensor open and connected gives error
- #9454: Object reference error when switching tabs in Configuration window
- #9460: Closing Virtual Forward Power Meter window shows error: Power Meter Init count: 1
- #9465: Carrier frequency control is not correctly sized on the VI Signal generator window
- #9496: Use default graph customization disables controls
- #9504: Report Generator: date format wrong
- #9507: Input string was not in a correct format.
- #9523: Keyword PEAK TABLE does not work in 2018.1.4
- #9533: Power Burst graphs are not correctly shown
- #9535: Configuring a multiband test has a slow response
- #9567: Table is turned after last peak measurement
- #9600: Object reference: Invoke RaiseProgressChangedEvent
- #9634: Isotropy positioner is not being initialized at first angle at the begin of the test.
- #9635: Sudden shutdown without warning.
- #9637: The question 'Are you sure you want to close the test' is shown twice.
- #9638: Opening a small amplifier linearity test result is taking minutes
- #9639: Measure time and step frequency controls are not enabled correctly in RE Multiband when switching testsites
- #9653: Driver: CDN - MEB M5/25A
- #9657: App. Note 115: How to measuring Amplifier harmonics using RadiMation.
- #9658: RM 2018.1.3 crashes often, with object reference not set
- #9687: Multiband EFT is very difficult to close when the automated test is finished
- #9694: External Modulation options not always shown
- #9714: Deadlocking in clsDataIdentifiers.
- #9715: Move during measurement is not shown
- #9719: Polar plots don't scale correct.
- #9731: New sweep in a Multiband Emission test is not overwriting the previous sweep
- #9732: Testdata of a singleband immunity test is not saved at the end of a test
- #9734: Index out of bounds when shift clicking on a band.
- #9742: Changing a sequence file does not work correctly.
- #9753: Driver: Rohde & Schwarz switch matrix OSP230 Frame F01 with options OSP-B133, OSP-B104
- #9795: Object reference not set to an instance of an object.
- #9819: Fix all warnings in the RADDOCS project
- #9820: Fix all warnings in the RADTESTS project
- #9821: Fix all warnings in the RADIMAT project
- #9822: Fix all warnings in the RADEMITESTS project
- #9823: Fix all warnings in the INTERPOLATIONCHART project
- #9824: Fix all warnings in the SAVEUNICODEFILETESTS project
- #9825: Fix all warnings in the INTERPOLATIONTESTS project
- #9826: Fix all warnings in the RADSTDCLTESTS project
- #9827: Fix all warnings in the HTTP_POST project
- #9838: Crash during report generation
- #9855: Set all C# projects to 'Treat all warnings as error' to 'All'
- #9874: This is not a valid action for the end of a row.
- #9877: Radiated emission bandwidth is not correctly displayed in 2019.1.1
- #9884: EUT window not visible
- #9891: Table is not visible during test
- #9897: Too much limitlines in MB RE is causing radimation to get slow and hang
- #9922: Driver: AR250S1G6M1 Amplifier
- #9923: Driver: Coupler, Amplifier Research: DC7230A
- #9924: Driver: Amplifier AR5000A225A
- #9961: Incorrect bandwidth in report for CE and RE
- #10019: IF Overload is detected ESU26.
RadiMation version 2018.2.9[edit]
- #9094: Cable Current sensor => powermeter is not taken into account during Conducted Immunity Singleband tests.
- #9787: Maturo NCD The connection for the given session has been lost.
- #9917: Der Objektverweis wurde nicht auf eine Objektinstanz festgelegt.
- #9939: Maturo NCD De-Init Freeze
RadiMation version 2018.2.8[edit]
- #8151: Spikes in CE MB measurement HAMEG HMS3010
- #8932: Incorrect measurement results Conducted Immunity substitution
- #9434: Final measurement is performed on another location than the optimized location
- #9911: Generating a report is not working.
- #9936: CANalyzer 9 SP5 using CAPL via Canoe switch-matrix validated for windows 10
- #9954: How to perform a closed loop method test accoording ISO 11452-4
RadiMation version 2018.2.7[edit]
- #5562: Adjust the calculation of the Bursts to be in compliance with the 1.8.2 version of EN 300 328
- #6732: Calculate TxSequence and TxGap times accoording to ETSI EN300328 V1.9.1
- #9298: Duty Cycle of Burst measurement is not correctly calculated
- #9689: Height graph in Multiband emission is showing up to 2.4 km.
- #9727: Driver: BONN BAS 0060-4
- #9752: Driver: Rohde & Schwarz signal generator SMB100B
- #9789: Antenna height does not scale properly.
RadiMation version 2018.2.6[edit]
- #9177: Graphline is not drawn in radiated immunity test
- #9770: ErrorReport: "Could not load file or assembly 'MindFusion.Graphs, Version=2.3.2.451"
- #9771: Strange 'CheckButton 1' button in virtual instrument window of RPR3006W
- #9785: Driver: Werlatone C3908
- #9896: R&S ESR26: Specific system error -200 : "Executio error: function not available. SCAN1:Band:RES 120 Khz
- #9913: Explain how to use the interface between RadiMation and CANoe
RadiMation version 2018.2.5[edit]
- #8751: Renaming a test-level input in a multiband immunity test, removes the second, if they both have the same name.
- #9055: Remote operation of the RadiLink is not always indicating error situations
- #9324: ErrorReport: "Collection was modified; enumeration operation may not execute.
- #9335: VI Window crash at closing
- #9517: Keywords changed
- #9591: Object reference not set to an instance of an object: ctlControlStop.CancelCommands(
- #9604: Report Error does not handle polish characters
- #9670: Auto stepsize of TDEMI is set to 5 kHz, even if 120 kHz RBW is selected
- #9685: Attachments of the EUT File are corrupted
- #9693: Add printout templates for CI Callibration and C Susceptibility
- #9696: Calculated voltage graph keyword is now working
- #9714: Deadlocking in clsDataIdentifiers.
- #9717: Change menu order for Pulsed Imunnity tests.
- #9742: Changing a sequence file does not work correctly.
RadiMation version 2018.2.4[edit]
- #8381: Impossible to remove graph line from different test data when the file has been closed
- #9572: Red dots as quasi peak markers
RadiMation version 2018.2.3[edit]
- #7975: Communication interrupt during EFT test
- #8041: Agilent 9038A as powermeter
- #8221: RadiMation logo in graph is not shown first time a test is opened
- #8280: Test equipment for EUT Calibration of Attenuation / Gain cannot be changed
- #9679: Cannot find turntable in the currently active testsite.
RadiMation version 2018.2.2[edit]
- #9595: R&S ESI 40 GPIB randomly stops communication
RadiMation version 2018.2.1[edit]
- #8263: Opening non existing file with RadiLog crash Gives exception
- #9002: Also show the latest DRIVERS.EXE version on the RadiWiki Main page
- #9115: Temp filenames in creating new files
- #9202: Adding input fieldprobe without enabling any probe in RI MB test, is causing object reference
- #9224: In ErrorReport an unknown label 'usr agr' is present
- #9246: Script for copying latest RadiMation release to Radi-Products Production folder
- #9300: How to show the points only that have the status 'fail'
- #9301: ErrorReport: "ETS-Lindgren EMField 7008-005 FWD power: Parameter too low"
- #9312: EMI reciver switch to spectrum analyzermode while initialisation for conducted emission testing
- #9323: SystemCrash: Source array was not long enough. Check srcIndex and length, and the array's lower bounds.
- #9333: ErrorReport: "Cannot find the device driver.
- #9334: ErrorReport: "Startfrequenz ist zu niedrig fuer das gewaehlte Geraet"
- #9340: Driver: Shaffner Broadband Horn Antenna 1-18 GHZ - BHA 9118
- #9358: Radimation crashes when device is in use by other software
- #9372: How to remote control LISN R&S ENV216 with the R&S ESW 26 via radimation
- #9373: Rohde & Schwarz SMR 20 (ID7340): Specific System Error: 110,"Output unleveled"
- #9389: How do we set-up for testing acording EN6100-3-2 and EN 61000-3-3
- #9390: The file extensions of the template and the output document do not match
RadiMation version 2018.1.8[edit]
- #9790: Conducted Emission measurement with R&S ESPI 3
- #9846: Extra peak measurement after CE measurement
RadiMation version 2018.1.7[edit]
- #9956: Radiated Immunity MB - Graph is not drawn
RadiMation version 2018.1.6[edit]
- #5636: R&S ESPI 3 error Data out of range SCAN1:BAND:RES
- #6351: How to configure the Monitoring input channel setup for the used AD Converter device driver
- #6377: How to enable or customize the borders of the grid of the exported table information of the ||EQUIPMENT TABLE|| keyword
- #6380: Would it be better to do fewer sweeps with a long sweep time or many sweeps with short sweep time?
- #6392: How is the Monitoring input channel formula option used?
- #8434: Dongle Failure ZF Friedrichshafen AG
- #8495: Not checked before Power delivery check if the RadiField amplifier is still in operate
- #8547: Some report generator codes no longer work
- #8751: Renaming a test-level input in a multiband immunity test, removes the second, if they both have the same name.
- #8760: The CheckDevice check for Amplifier Research amplifiers using the P0/P1 command set is always succesful
- #8795: Driver: switch matrix: MCC USB-ERB24
- #8797: Driver: Antenna tower controller HD GmbH IEEE488-16REL GPIB
- #9055: Remote operation of the RadiLink is not always indicating error situations
- #9142: Driver: Mini-Circuits PWR-6GHS
- #9203: RadiSense probe returns unknown ERROR 101
- #9331: Radiated immunity RF on but not scanning
- #9547: Schloeder CWG500 is waiting forever on a discharge trigger
- #9551: Problem with ECU-6 Communication
- #9559: RF Amplifier (ID1344): Prana returned a warning state.
- #9575: Problems controlling the Keysight Technologies CXA (N9000B)
- #9590: Index was outside the bounds of the array. clsDataIdentifiers.Add
- #9591: Object reference not set to an instance of an object: ctlControlStop.CancelCommands(
- #9604: Report Error does not handle polish characters
- #9610: Driver: Amplifier VBA250-80
- #9611: Driver: Clamp Schwarzbeck CVP 9222B
- #9612: Driver: CDN FCC-801-M2-16
- #9613: Driver: LISN - Rohde & Schwarz: ENY 41
- #9614: Driver: Pulse Limiter - Rohde & Schwarz ESH3-Z2
- #9615: Driver: LISN - Schwarzbeck CAT3 8158
- #9617: Driver: Calibration Adapter FCC-801-150-50
- #9632: Spectrum Analyser (ID1691) Input 2: Specific System Error: -420,"Query UNTERMINATED; FREQ:STOP?
- #9644: Antenna and turntable do not return to zero
- #9645: R&S ESCI 7 - RBW switches to auto 3 MHz
- #9647: Closing a device window crashes radimation - Exception has been thrown by the target of an invocation
- #9673: Collection was modified; enumeration operation may not execute.
- #9685: Attachments of the EUT File are corrupted
- #9701: Driver: Keysight Technologies MXA N9020B + CXA N9000B
- #9709: Driver: Schaffner SCR 3501
- #9713: Communication time out RPR2006C
- #9754: test interrupted with "Cancel" and resumed with "Automatic" with wrong RBW and wrong traces
- #9761: ErrorReport: "RBW switched to Auto(100kHz) during 30M-1G test
- #9775: Rohde & Schwarz ESPI 3 Scanning Receiver: Specific System Error: -222,"Data out of range; SCAN1:BAND:RES 9 KHZ.
- #9783: Driver: MT 400 D
- #9798: When RE MB test is interrupted and resumed, another RBW is being used
- #9847: RBW changed, radimation halted
RadiMation version 2018.1.5[edit]
- #2338: Driver: R&S SWM with external modulation
- #5636: R&S ESPI 3 error Data out of range SCAN1:BAND:RES
- #8066: Extend the DSA832 pre-amp option detection
- #8113: Unable to communicate with CWS500N 1.2
- #8748: Use configurable driver for R&S UPV EUT monitoring
- #8514: Driver: emtest Compact NX
- #8521: Detection if the Keysight GPIB card is used is not working anymore with Keysight IO Libraries Suite 2018
- #8715: Rigol DSA 832: Specific System Error: -220, "Parameter error"
- #8796: Driver: Luthi CDN's 801-M1/M2/M3
- #9139: Driver: Denkovi DAE-CB/Ro16-JQC3FC-USB
- #9178: Unknow error - Change text when wiki can not find a page
- #9316: Software hangs in "detecting peak 100%
- #9438: Not all data from PMM9010 is received due to RS232 buffer overrun
- #9379: Graphline does not display (RBW 10 Hz)
- #9400: Driver: DARE!! Instruments DCN CDN's
- #9405: Report Generator changed font size in table causes errors
- #9416: Communication with R&S ESPI3
- #9423: Driver: R&S FPC1000
- #9455: Singleband frequency rounded
- #9457: Driver: Signal generator Aeroflex 2023
- #9463: No validation error is shown on the erroreport dialog in a new installation
- #9464: R&S SML signal generator should not use RSSML::msBuffer class member
- #9497: RPR2006C Wrong Command
- #9505: The red dashed 5.1 and 3.1 dB limit lines are not shown in the amplifier linearity test
- #9516: PMM9010: Incorrect start and-or stop frequency when 1 MHz RBW is used
- #9517: Keywords changed
- #9522: Amplifier Reflected power does not show 2018.1.4
- #9530: emtest PFS 500N: One or several values are limited by the simulator. The parameters are set to their minimum.
- #9537: Backport logging in clsGraphDrawer.Update to 2018.1.5
- #9543: Could not load file or assembly 'MindFusion.Common.WinForms'
- #9565: Radifield 2018 Error 1 on init. Wrong command
- #9573: Rigol Technologies DSA 815- Preamplifier option is not available in this device. Select no preamplification (0 dB)..log
- #9576: Destination array was not long enough
- #9578: GPIB Keysight USB GPIB 82357B Connection Expert 2018
- #9579: RadiSwitch Error
- #9581: RSS2010S
- #9589: Source array was not long enough. clsGraphElements.Add
- #9596: Report Generator table, generates multiple columns for 1 column name.
- #9598: Rigol DSA815: Unexpected RBW bandwidth
- #9606: Maturo Controller MCU N933 @7: Device stream is not connected.
- #9616: Not all spectrum analyser device drivers return the correct RBW mode during the GetRBW call
- #9622: "GAUSS INSTRUMENTS TDEMI X18: Insufficient location information or the device or resource is not present in the system. "
- #9642: EM Test NX 5:Change default port number to 1234
- #9643: Radimation crashes after error: Can not send/save the error report
- #9648: GAUSS INSTRUMENTS TDEMI X18: Specific System Error: 1
- #9649: Gauss Instruments TDEM X18: Ethernet not connected
- #9654: Driver: Receiver / Spectrum analyser Hewlett Packard 8591EM
- #9655: Driver: Signal generator - Marconi instruments 2023
- #9664: Agilent E8257D-567 (without ALC): Specific System Error: -221,"Settings conflict; Pulse period set less than pulse width value. Pulse width changed to match period value
- #9666: GAUSS INSTRUMENTS TDEMI X18: Specific System Error: -114,"Header Suffix Out of Range;SENS:SCAN1:TIME?"
- #9668: Source array was not long enough. Check srcIndex and length, and the array's lower bounds.
- #9672: Setting the BIAS Target of the RadiLink is not working
- #9675: Changing unit for Attenuation/Gain test does not work
- #9677: Unexpected power delivery test failed
- #9678: Power delivery test failed
- #9756: RBW switches from 120 kHz to Auto(1 MHz) during test with 2108.1.5 and R&S ESCI 7
RadiMation version 2018.1.4[edit]
- #5261: Frequency plot can't be printed
- #5636: R&S ESPI 3 error Data out of range SCAN1:BAND:RES
- #5783: R&S ESU 26: Error -222: Data out of range. Depending on the preamplifier that is used
- #6373: R&S ESU 40: Error -222: Data out of range SCAN1:BAND:RES 1 KHZ
- #6811: Add new RadiMation startup Reseller logo
- #7329: Using multiple bands that switch preamplifier on and off gives error: R&S Receiver ESU 8
- #7482: RBW and SpanCentreFrequency order. Unable to set frequency multiband
- #7728: Unicode: clsFile EUT information is saved as ascii. Unicode characters are saved and read as '???'
- #8590: Unknown error when Visa is not installed, and is tried to control the EMCenter by LAN
- #8600: Spectrum Analyser frequency out of range with PreAmplifier MB
- #8691: Frequency out of range at second measurement.
- #8730: Source array was not long enough. Check srcIndex and length, and the array's lower bounds.
- #8734: controlling radigen over windows rs232 to com3 over usb is not working
- #9129: RadiMation crash when GPIB-Board is not installed or not configured correctly (GPIB-EDVR)
- #9177: Unable to open measurement files (_Object reference not set to an instance of an object..)
- #9188: Corrupt testfile, no data is visible in the graph
- #9221: Receiver (ID1691) Input 2: Specific System Error: -222,"Data out of range; SCAN1:BAND:RES 1 KHZ"
- #9247: Wrong output for Keyword ||DESCRIPTION||
- #9274: Driver: Field probe Narda EMC-30 RS232
- #9275: Driver: Passive drivers Philips Saeco S.p.A.
- #9277: Driver: Passive drivers CDVI Wireless S.p.A.
- #9278: MB test with Carad as eut controller gives an error
- #9287: Test data disappeared
- #9293: Radimation restarts during emission measurements (read/write outside the bounds of array)
- #9310: Testdata again not visible
- #9311: Test data lost / not visible
- #9314: No response turn table
- #9335: VI Window crash at closing
- #9338: Driver: Shaffner Broadband Horn Antenna 1-18 GHZ
- #9339: Driver: Stanford Research Systems - Model DS335
- #9341: Driver: Agilent N1913A
- #9345: De index ligt buiten de matrixgrenzen - Index out of range
- #9350: Signal power went too high (-6 dBm) during power delivery test
- #9363: Driver: CDN LAPLACE S46 ST8
- #9364: Driver: LISN LAPLACE (Cranage) LISN16A1P
- #9365: Driver: Antenna Laplace SA1020/RF200 No remote control
- #9366: Driver: Injection device Teseq KEMZ 801
- #9367: Driver: Calibration JIG Teseq KEMZ 801
- #9368: Driver: CDN MEB M3, M2, AF3-1 and AF2-1
- #9369: Driver: Antenna ARA LPB-2513
- #9382: Unable to ReportGenerate Emission and Peak Table
- #9394: Index was outside the bounds of the array.
- #9396: Index was outside the bounds
- #9398: Driver: Bonn BLMA 2060-30
- #9409: R&S ESI does not show graph of logarithmic stepped sweep
- #9426: Subject of ErrorReport should contain the description of the error
- #9428: Transmitting an ErrorReport is not working, complains that it is not possible to send ErrorReport
- #9440: Word locks documents after that it is not possible to print
- #9443: ErrorReport: "De bronmatrix is niet lang genoeg. Controleer srcIndex en de lengte ervan, en de ondergrenzen van de matrix. ".
- #9521: Peak table does not work in 2018 version
- #9524: Object reference not set to an instance of an object: clsCommandSetTurnTableAngle.IManualModeAsyncCommand_StartWork(
- #9525: Antenna Tower (ID1365) Aut.: Impossible to get description string for this Error! Probably an unknown RadiMation Error. Code: 805308669 (0x300008FD))
- #9538: Version Information is not displaying
- #9539: Text does not display on the devices tab in 2018.1.4
- #9549: Problem exporting conducted emission
- #9558: ErrorReport: "Access to the path 'C:\Users\u0058578a\Documents\My Music' is denied. "
- #9577: DARERSS2010I
- #9584: Question: What is AT. Is this used for low frequent rod or loop antenna?
- #9593: Vertical stripes QP measurements
- #9608: Unhandled exceptions
RadiMation version 2018.1.3[edit]
- #3570: Turntable Frankonia FC02 not working correctly
- #5984: Printing results of a burst test a second time, prints the same data as the first time
- #7219: Radifield powermeter is not usable on 800 MHz
- #7977: Holaday Instruments HI-4422: Invalid command
- #8014: Extend the maximum size of a generated Report
- #8031: Default graphs during Multiband Immunity are not shown during test
- #8162: Radimation startup crash Recursive customization file
- #8379: Object reference error in updating the Emission Graph.
- #8537: SMR40 total crash of communication
- #8585: Radiated Emission Magnetic field goes right into manual mode
- #8588: Field test level keyword missing.
- #8609: Agilent 34401 multimeter is always selecting DC coupling in CS101 calibration and CS101 test
- #8912: Object reference clsIDocumentBase.UpdateAllViews update graph
- #9112: No match was found in the specified regular expression
- #9116: Fax number on about screen is not valid anymore
- #9119: Sharing Monitoring AD channels between tests, and test setups.
- #9141: Radimation crashes, no error report
- #9173: MXE ERROR -224: Settings conflict; only active antenna unit available; no other y axis units
- #9181: File > New > EUT is showing a strange filename
- #9195: RF Amplifier (ID1647); Device not connected
- #9253: Add the name and identifier of the Reseller as it is programmed in the dongle in the ErrorReport
- #9254: Add the identifier of the dongle in the ErrorReport
- #9263: Unhandled exception handler is crashing software before report can be transmitted
- #9271: Radimation stops: software tries to change RBW on R&S ESU 8
- #9272: RadiMation keeps waiting: RBW value changed by R&S ESU 8
- #9282: Problem using the EMS RSC controller
- #9283: Cannot Find the device driver (RPR2006P)
- #9285: R&S ESU 8 Input 2 Scanning Receiver: Specific System Error: -200,"Execution error;Function not available;ABORT"
- #9302: BLMA 1040-100D
- #9306: Incorrect table export of polar plot
- #9307: No match was found in the specified regular expression
- #9309: ErrorReport: "Rohde & Schwarz ESR 7 Input 1 Scanning Receiver: Specific System Error: -200,"Execution error; Function not available; ABORT""
- #9343: Driver for: Teseq CBA 16-150
- #9347: ErrorReport: "Fixed power test. Vrijdag Einde vd dag
- #9351: ReportError("Exception has been thrown by the target of an invocation.")
- #9352: Cannot install CTR 1001S in win 10
- #9370: ErrorReport: "Cannot find the device driver.
- #9509: Install issues due to download problems
- #9534: Decimal sign changes after radimation error
- #9546: Signal power went too high (-4 dBm) during power delivery test
- #9550: Keywords not woking in Radimation_Validation_Template v 4.2
- #9553: EN61000-4-6 Callibration and testing with two powermeters
- #9557: Rohde & Schwarz ESPI 3: Specific System Error: -200,"Execution error; Function not available; CALC:UNIT:POW DBUV
- #9560: ErrorReport: "File 'J:\postvak\testsite\Referentie metingen\Radiated Immunity Semi Anechoic Chamber EN61000-4-3\ID1494 - Cavendish\RI 80-1000MHz 10Vm AVG 2ndQ 2011.BMP' not found during a 'ShellExecute' call "
- #9588: Missing/empty cells in peak table
- #9594: Error .csv dump LeCroy Wavesurfer
- #9660: TSF file name changed.
- #9667: Not able to connect to devices in RadiCentre
RadiMation version 2018.1.2[edit]
- #4209: No frequency verification on correction file
- #8718: HP3478A Multimeter wrong mode selection.
- #8986: No graphline is shown anymore in the polar plot of a peak
- #9136: Driver: HP 85420E Pre-amplifier
- #9145: Driver: Aaronia antennas
- #9245: DARE!! Instruments RPR2006C--FWD: No match was found in the specified regular expression.
- #9270: Using R&S Z91 as monitoring power sensor with attenuation (-30 dB) takes a long time to measure
- #9403: RI: Certain graph lines are not added to the report. Outputing a table does work.
- #9663: Radisense 10 RSS2010B/H cor. OFF: I2C Not-Acknowledge (NACK) "
RadiMation version 2018.1.1[edit]
- #8567: Include Sentinel System Driver 7.6.0 in setup program
- #8720: Power is sometimes regulated below the desired testlevel of 1 watt
- #8773: Configurable RS232 stream always appends \n after command
- #8865: PDF icon for the manual on the desktop is always using a white icon. Not showing the PDF icon
Known Problems[edit]
R&S spectrum analysers/receivers communication problems[edit]
It has been detected that on very fast PC’s, the communication with modern Rohde & Schwarz spectrum analyser/receivers can be corrupted. These problems are related to the speed of the PC and the used GPIB bus. The R&S receivers that are related to this problem are: ESI/ESIB/ESPI/ESCI/FSE/ESU.
We strongly advise that you slow down the GPIB communication speed between the PC and R&S receivers. Follow these steps to do so:
- Open the configuration window of RadiMation®
- Select the Device Drivers tab
- Select Receivers/Spectrum Analysers
- Select the R&S receiver
- Press Edit to open the device driver configuration window
- Press the Advanced button
- The GPIB configuration dialog of the R&S device driver will be shown
- Activate the Show advanced Settings checkbox
- The GPIB Configuration dialog will show a Wait time box
- Specify 3000 us in the Wait time box
- Close all windows, and repeat these steps for each and every R&S receiver.
Configuring the R&S device drivers in this way will slow down the GPIB bus, by ensuring that at least 3 milliseconds delay are included between each GPIB call to the R&S devices
Previous RadiMation® version raises a run-time error[edit]
If the latest RadiMation® version and an older RadiMation® version are installed on the same PC, it can happen that the older RadiMation® version raises an error during the start of RadiMation.
This error is raised because a file that is shared by both versions is written in a newer file format that cannot be read anymore by older software. The solution for this problem is to remove the file: C:\Program Files\DARE Development\RadiMation\Confdvdr\CUSTOMIZ.DAT. The CUSTOMIZ.DAT file is used to remember customizations of the colour of the graphs, so removing the file will not influence the measurements. However the graph colours will probably be changed to the defaults as specified by RadiMation®.
Windows 8, Windows 7, XP, Vista and older are not supported[edit]
Due to the fact that Microsoft has ended their support for these versions we are also forced to stop the support of RadiMation® running on these operating systems. This means that we do not test if RadiMation® is working correctly on these operating systems. However it could be that RadiMation® is still working correctly. We strongly suggest upgrading your operating system to Windows 10 (21H2). Both the 32-bit and the 64-bit versions of the Microsoft Windows Operating system are supported.