== 1.3 Configure the data readout of the device ==  
 
== 1.3 Configure the data readout of the device ==  
  
{{ScreenElementDescription|Trigger and Reading|
 
  
{{ScreenElementDescription|Trigger|The command need to trigger the measurement. A common used [[SCPI]] is *TRG. When left blank no command will be send.}}
+
{{ScreenElementDescription|Trigger|A trigger command can be specified to request a measurement from the device. A common used [[SCPI]] is *TRG. When left blank no command will be send.}}
  
{{ScreenElementDescription|Read Back|In the read back the returned information is defined. This is different for every type of machine there for {{radimation}} used [[regular expressions]] to understand the format. A good regular expression for finding the first number in scientific notation in a text is:
+
{{ScreenElementDescription|Read Back|Different devices return different messages with measurement values. The configurable AD convertor only need the numeric value of the measurement message. There for, a [[regular expressions]] can be set in the Read Back.
 +
A good regular expression for finding the first number in scientific notation in a text is:
  
 
  '''([+-]?[0-9]*\.?[0-9]+([eE][+-]?[0-9]+)?)'''
 
  '''([+-]?[0-9]*\.?[0-9]+([eE][+-]?[0-9]+)?)'''
  
For more possibilities see the example section  
+
For more possibilities see the example section.
}}
 
 
}}
 
}}
  
{{ScreenElementDescription|Trigger and Reading|
+
==== Minimum and maximum value ====
 +
Minimum and maximum can be set to limit the measured value. And to calculate the measured value to a specific unit.
  
 
{{ScreenElementDescription|Minimum value|The lowest value that can be measured with this device. The value must be the same as filled in the [[EUT]] window.}}
 
{{ScreenElementDescription|Minimum value|The lowest value that can be measured with this device. The value must be the same as filled in the [[EUT]] window.}}
  
 
{{ScreenElementDescription|Maximum value|The highest value that can be measured with this device. The value must be the same as filled in the [[EUT]] window.}}
 
{{ScreenElementDescription|Maximum value|The highest value that can be measured with this device. The value must be the same as filled in the [[EUT]] window.}}
 
}}
 
 
== 1.4 Configure multiple channels for 1 device ==
 
  
  
== 1.5 Examples ==
+
== 1.4 Examples ==
  
 
{| class="Regex example table"
 
{| class="Regex example table"

Revision as of 12:50, 6 October 2016

How to measure data from the EUT using the Configurable AD Convertor device driver

This Application Note explains how a configurable AD convertor can be used to monitor an EUT. With the AD convertor, any type of signal can be monitored and plotted within RadiMation during a test.

  1. First the AD convertor must be configured so values can be read within RadiMation.
  2. Second the value must be plotted using the correct unit types.


1 Configuring the AD convertor

Configurable AD Converter Configuration Window.png

1.1 Specify the communication type

First the communication type must be configured for the AD convertor.

ConfigureCommunication.png

1.2 Configure the initialization of the device

ScreenElementDescription.png Reset The reset code that needs to be send to device. When left blank no command will be send.
ScreenElementDescription.png Init The Init code that needs to be send to device. When left blank no command will be send.
ScreenElementDescription.png Get Id The code that needs to be send to device to get the identification back. A common used SCPI command is *IDN?. When left blank no command will be send.



1.3 Configure the data readout of the device

ScreenElementDescription.png Trigger A trigger command can be specified to request a measurement from the device. A common used SCPI is *TRG. When left blank no command will be send.
ScreenElementDescription.png Read Back Different devices return different messages with measurement values. The configurable AD convertor only need the numeric value of the measurement message. There for, a regular expressions can be set in the Read Back.

A good regular expression for finding the first number in scientific notation in a text is:

([+-]?[0-9]*\.?[0-9]+([eE][+-]?[0-9]+)?)

For more possibilities see the example section.

Minimum and maximum value

Minimum and maximum can be set to limit the measured value. And to calculate the measured value to a specific unit.

ScreenElementDescription.png Minimum value The lowest value that can be measured with this device. The value must be the same as filled in the EUT window.
ScreenElementDescription.png Maximum value The highest value that can be measured with this device. The value must be the same as filled in the EUT window.


1.4 Examples

Received information
Regex
RadiMation Readout
Note
U_L_N 325
(-?[0-9.,Ee-]+)
325
Takes the first number.
I_L1;12.34
;(-?[0-9.,Ee-]+)
12.34
Takes the first number after ";".
THD_U_L1;;14,5
;;(-?[0-9.,Ee-]+)
14.5
Takes the first number after ";;".
AC_FREQ;Channel1;1.23E3
;*;(-?[0-9.,Ee-]+)
1230
Takes the first number after the appearance of a second ";".


Read backs can always be tested trough the "Validate Expression tool"

ValidateExpressionTool.png<\div>